Ophir-Spiricon IBP-YAG Series User Manual Download Page 12

Two setup files have been provided with each IBP-YAG measurement head.  The setup files 

contain calibration information and are unique to each IBP-YAG head.  The setup files are 

associated with the serial number of the IBP-YAG hardware in the form:  

~YAG<Serial number>_CW.cfg   or  

~YAG<Serial number>_PULSED.cfg 

 
For the initial checkout, select the ‘CW’ setup file.  If you have more than one measurement 

head, make sure that you select the file with the serial number matching the IBP hardware. 
To select the configuration file click:  FileRestore Config…, and the …CW.cfg file shown. 
 

2.3.4

 

Verify that the system is operating. 

Click Start! If everything is working correctly LBA-PC will automatically start collecting data from 

the detector and displaying it on the screen.  Since we aren’t using a laser yet, the Beam display 

window will be dark.   

The first indications you will have that frames are being accepted from the detector are 

the displayed results on the left side of the screen and the frame counter in the bottom 

right hand corner of the screen.  If the numbers in the results are changing and the 

frame counter is incrementing, then LBA-PC is accepting frames from the detector, even 

though no beam is shown on the display. 
If the software does not appear to be collecting frames, verify that LBA-PC is operating in 

CW mode.  Click Stop!. Open the Options, Capture… dialog box and under Method 

select the Continuous option. Then click OK.   
If there still is no indication of operation, verify in the lower status bar that the Video “V” 

enunciator is illuminated Green.  If this indicator is red, it is an indication that the frame 

grabber is not seeing an operating camera. In this case, refer to the troubleshooting 
sections of the software manual or contact the Spiricon Service Dept. 

2.4

 

Prepare the laser for collecting data 

Follow these steps to prepare the laser for taking measurements. 

 

Verify that the operating settings for the laser will cause the power levels and beam size to fall 
within the ranges indicated in Appendix B.  You may want to use a lower output power while 

aligning the IBP-YAG to the laser 

 

Remove any focusing optics from the delivery system of the laser. 

The IBP-YAG  is designed to operate with collimated lasers 

only.  Using a focused Laser beam with the IBP-YAG will 

damage the instrument. 

 

If the laser is equipped with assist gas, temporarily disable the gas delivery. 

 

The computer system must be set up close enough to the laser so that the cables will reach to 
the IBP-YAG.  Place the IBP-YAG on the work surface with the input aperture under the beam 

delivery tube.  Route the hoses and cables out of the way.  Make sure that the controls on the 

back of the IBP-YAG are accessible. 

 

Level the IBP-YAG using the leveling adjusters and bubble level.  Center the input aperture of the 
IBP-YAG in the beam path.  If the laser is equipped with a visible aiming beam, use it on the 

IBP-YAG 

Setup and Operation 

2-3 

Summary of Contents for IBP-YAG Series

Page 1: ...W 1000 N Main Logan UT 84321 For Sales Service or Technical Support Phone 435 753 3729 Fax 435 753 5231 Email service ophir spiricon com Email sales ophir spiricon com Copyright 2008 Ophir Spiricon In...

Page 2: ...manual at any time and without notice All rights to the product and any accompanying operators manuals are reserved While every precaution has been taken in the preparation of this product the publis...

Page 3: ...__________ 2 1 2 3 Frame Grabber Setup ________________________________________2 2 2 3 1 Refer to your LBA PC Operators Manual ___________________________ 2 2 2 3 2 Starting LBA PC for the first time_...

Page 4: ...and use procedures We also suggest that you read the software manual to become familiar with the controls and the measurements provided Chapter 1 contains general Information and specifications of the...

Page 5: ...ndensing Note When storage temperature may drop below 5 C all cooling water must be removed from the unit to prevent damage due to freezing Operating Temperature 10o C to 50o C 50 F to 122 F Operating...

Page 6: ...ons Weight 12 kg 26 lb Dimensions 24 x 50 x 32 cm 9 6 x 19 5 x 12 6 17 8cm 7 0 32 1cm 12 6 22 8cm 9 0 26 0cm 10 2 48 4cm 19 1 24 3cm 9 6 19 7cm 7 8 45 7cm 18 0 Figure 1 1 Overall Dimensions IBP YAG Ge...

Page 7: ...to connect the IBP YAG to the LBA 7xx frame grabber card 3 Cooling Water Inlet 4 Cooling Water Outlet These fittings are standard barbed hose connectors for 5 16 ID Hose Refer to appendix A for water...

Page 8: ...e shutter closed except when the laser is directed into this opening 9 Shutter Release Lift up on the top handle to close the shutter 10 Bubble Level 11 Leveling Adjusters Use the leveling adjusters t...

Page 9: ...tive eyewear when operating the laser Do not set up the IBP YAG or any optical components at eye level Do not wear reflective jewelry or watches while operating this equipment to reduce the possibilit...

Page 10: ...nal Cart system If ordered the cart system will be shipped separately and will include separate unpacking and setup instructions 2 2 Setting up the IBP YAG 2 2 1 Frame Grabber and Software Setup The I...

Page 11: ...this connection remove the plug from the back of the instrument and replace it with an 1 8 NPT hose fitting Connect to a supply of clean dry nitrogen or air at about 5 psi 2 2 2 3 Power and Video conn...

Page 12: ...dialog box and under Method select the Continuous option Then click OK If there still is no indication of operation verify in the lower status bar that the Video V enunciator is illuminated Green If t...

Page 13: ...tions or when using chilled cooling water run the purge air for 10 min before starting the cooling water Set fine attenuation knob to the minimum position by turning the knob clockwise until it stops...

Page 14: ...saturated portions of the displayed beam will turn white when saturated Increase the fine attenuation until beam is just below saturation Center the Beam image in the display by adjusting the levelin...

Page 15: ...r s shutter or by turning the laser off Never attempt to block the laser beam by closing the IBP YAG shutter The IBP YAG shutter is not designed to handle high energy beams and doing so will damage th...

Page 16: ...cs 3 2 1 Cleaning the sampling optics To remove small amounts of dust from the primary sampling mirror open the shutter and use photographic type canned air and lens brushes to carefully clean the sam...

Page 17: ...acetone or isopropanol onto the lens tissue wetting the entire surface of the optic Without lifting the lens tissue drag the tissue across the optic just fast enough so that the solvent evaporates beh...

Page 18: ...aning or replacement Remove the Leveling Plate to allow access to the bottom of the instrument The Leveling Plate may be removed by carefully prying down on the plate next to each of the leveling feet...

Page 19: ...instrument The shutter is attached to the front panel with 2 springs Do not let the shutter or springs touch the sampling mirror Figure 3 4 The Mirrors may be removed by removing 3 screws and clamps A...

Page 20: ...entify it is indicated with an arrow mark on the edge of the mirror Figure 3 5 Before reassembly make sure that all mechanical and optical surfaces are clean and free of dust Reassemble the instrument...

Page 21: ...o prevent water damage to the measurement head purge air must be provided DO NOT USE DE IONIZED DI WATER DI WATER WILL QUICKLY CAUSE SEVERE CORROSION OF ALL COMPONENTS THAT CONTACT THE WATER Water Pre...

Page 22: ...1 8 30 9 87 6 35 2 95 3 39 4 102 9 7 6 2 0 30 5 86 9 34 3 93 7 38 1 100 6 8 3 2 2 30 1 86 2 33 6 92 5 37 1 98 7 9 1 2 4 29 8 85 7 33 0 91 5 36 2 97 2 9 8 2 6 29 6 85 3 32 5 90 6 35 5 95 9 10 6 2 8 32...

Page 23: ...0 561 6 4 000 160 16 0 808 7 4 000 220 22 1 1 8 4 000 290 29 1 4 9 4 000 360 36 1 8 10 4 000 450 45 2 2 11 4 000 540 54 2 7 12 4 000 650 65 3 2 13 4 000 760 76 3 8 14 4 000 880 88 4 4 15 4 000 1 010...

Page 24: ...ower Energy Diameter 2 4 8 15 30 60 120 250 500 1000 2000 4000 Watts mm 0 03 0 06 0 13 0 25 0 5 1 2 4 8 16 32 64 125 250 500 1000 Joules 4 0 1 1 1 1 2 2 2 3 3 3 4 4 4 4 5 0 1 1 1 2 2 2 2 3 3 3 4 4 4 6...

Page 25: ...C 2 IBP YAG Laser settings Appendix C Calibration Information...

Reviews: