3
1
Dealer Maintenance Schedule
Perform the checks/actions listed in the Operator Maintenance Schedule as
well as those listed in the table below.
To avoid potential injury, use
United Care
parts only
when replacing parts.
To prevent the
user
from falling (or objects from falling)
that may cause injuries to the
user
, the operator or
other persons, always reinstall the track end stops if
they were removed for servicing. Note that the
recommended tightening torque is 20 N-m (15 lb-ft).
Figure 15
Dealer Check/Action
Frequency
Check the
hoist
casing for any damage
Every year
Check the
hoist
transmission
Every year
Check the
hoist
connecting joints
Every year
Check that the emergency brake is turning freely
Every year
Check that all emergency devices are working properly
Every year
Perform a load test on the
hoist
and track system with the
Safe Working Load
Every year
Replace the
hoist
strap
Every 2 years
Lubricate roller and spool shaft
Every 2 years
Lubricate transmission spool gear
Every 2 years