background image

 

HOBO Temp/RH 3.5% Data Logger (UX100-003) Manual 

1-800-LOGGERS 5 

www.onsetcomp.com 

Notes: 

 

Once the logger is launched, alarms will trip as 
determined by these settings. Logger alarms will display 
on the LCD screen. Note that the alarm limits are only 
checked when the logger’s LCD screen refreshes every 15 
seconds. 

 

The actual values for the high and low alarm limits are set 
to the closest value supported by the logger. For 
example, the closest value to 85°F that the UX100 series 
logger can record is 84.990°F and the closest value to 
32°F is 32.043°F. In addition, alarms can trip or clear 
when the sensor reading is within the logger 
specifications of 0.02°C resolution. This means the value 
that triggers the alarm may differ slightly than the value 
entered. For example, if the High Alarm is set to 
75.999°F, the alarm can trip when the sensor reading is 
75.994°F (which is within the 0.02°C resolution). 

 

When you read out the logger, high and low alarm levels 
will be displayed on the plot along with “Chan <#> Alarm 
Tripped” and “Chan <#> Alarm Cleared” events showing 
when the sensor alarm tripped and cleared. The “Chan 
<#> Alarm Cleared” event contains the value that was 
furthest out of range for the sensor before the alarm 
cleared (see the Points table for the actual value). 

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, let’s say the logger is recording data at a 5-minute 
logging interval and burst logging is configured to log every 10 
seconds when the temperature goes above 85°F (the high level) 
or falls below 32°F (the low level). This means the logger will 
record data every 5 minutes as long as the temperature 
remains between 85°F and 32°F. Once the temperature reaches 
90°F, for example, the logger will switch to the faster logging 
rate and record data every 10 seconds until the temperature 
falls back below the high level (or 85°F in this case). At that 
time, logging then resumes every 5 minutes at the normal 
logging interval. Similarly, if the temperature falls to 30°F, for 
example, then the logger would switch to burst logging mode 
again and record data every 10 seconds. Once the temperature 
rises back to 32°F, the logger will then return to normal mode, 
logging every 5 minutes. 

To set up burst logging: 

1.

 

Select Burst for Logging Mode in the Launch Logger 
window. If Burst has already been configured for this 
logger, click the Edit button in the Launch Logger window. 
Note that you cannot set up alarms with burst logging. You 
also cannot select the Stop Logging option “Never 
(wrapping)” if burst logging is configured. 

2.

 

In the Burst Logging window, select a sensor. In the 
example at right, the temperature sensor was selected. 

3.

 

Select the High Level checkbox if you want to set up a 
condition in which burst logging will occur when the sensor 
reading rises above the high level value. Type in the value or 
drag the red upper slider (see example at right). 

4.

 

Select the Low Level checkbox if you want to set up a 
condition in which burst logging will occur when the sensor 
reading falls below the low level value. Type in the value or 
drag the blue lower slider. 

 

5.

 

Repeat steps 2 through 4 for the other sensor, if desired. 

6.

 

Set the burst logging interval, which must be less than the 
logging interval. Select either a preset burst logging interval 
or select Custom and enter your own interval. Keep in mind 
that the more frequent the burst logging rate, the greater 
the impact on battery life and the shorter the logging 
duration. 

7.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging Mode 
in the Launch Logger window to make additional changes. 

Notes: 

 

Once the logger is launched, the high and low burst 
logging levels are only checked when the logger’s 
LCD screen refreshes once every 15 seconds. Therefore, 
if you set the logging interval to less than 15 seconds and 
the sensor reading falls outside the levels, the burst 
logging will not begin until the next 15-second refresh 
cycle. 

 

If high and/or low levels have been configured for more 
than one sensor, then burst logging will begin when any 
high or low condition goes out of range. Burst logging will 
not end until all conditions on all sensors are back within 
normal range. 

 

The actual values for the burst logging levels are set to 
the closest value supported by the logger. For example, 
the closest value to 85°F that the logger can record is 
84.990°F and the closest value to 32°F is 32.043°F. 

 

Burst logging mode can begin or end when the sensor 
reading is within the logger specifications of 0.02°C 
resolution. This means the value that triggers burst 
logging may differ slightly than the value entered. For 
example, if the High Level for a temperature alarm is set 
to 75.999°F, burst logging can start when the sensor 
reading is 75.994°F (which is within the 0.02°C 
resolution). 

 

Once the high or low condition clears, the logging 
interval time will be calculated using the last recorded 
data point in burst logging mode, not the last data point 
recorded in “normal mode.” For example, let’s assume 
the logger has a 10-minute logging interval and logged a 
data point at 9:05. Then, the high level was surpassed 

Summary of Contents for HOBO UX100-003

Page 1: ...ion 0 07 at 25 C 77 F and 30 RH Response Time 43 seconds to 90 in airflow of 1 m s 2 2 mph Drift 1 per year typical Logger Logger Operating Range Logging 20 to 70 C 4 to 158 F 0 to 95 RH non condensin...

Page 2: ...to connect the logger to the computer or the HOBO U Shuttle via USB cable see Setting up the Logger and Reading Out the Logger LCD Screen This logger is equipped with an LCD screen that displays detai...

Page 3: ...ng the launch configurations onto the logger from HOBOware Make sure the USB cable is connected to both the logger and the computer and try launching again The logger has been stopped with HOBOware or...

Page 4: ...button for 1 second if you select this option 10 Click the Start button to launch the logger Disconnect the logger from the computer and deploy it using the mounting materials see Mounting the Logger...

Page 5: ...burst logging You also cannot select the Stop Logging option Never wrapping if burst logging is configured 2 In the Burst Logging window select a sensor In the example at right the temperature sensor...

Page 6: ...you want to calculate during logging Note that Average is automatically enabled when selecting Standard Deviation Important Statistics apply to all enabled sensors every selected statistic will be cal...

Page 7: ...that the pins are in the two holes on the board shown here enlarged and without the logger case for illustrative purposes 4 Snap the louvered door back into place in the logger 5 Check logger status i...

Page 8: ...e lithium battery The battery may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case Do not dispose of the logger or battery in fire Do not ex...

Reviews: