16
Measuring Method
5.2 Assembling the measurement system
!
CAUTION
•
Ono Sokki DF-2200 On-Board Flow Meter/FM Series Digital Flowmeter and the flow signal cable (FP-0015/0016/0017) and
the temperature signal cable (FP-0025/0026/0027) are sold separately. Purchase the appropriate product according to your
usage and purpose.
For details, contact the nearest Ono Sokki sales office or the distributor where you purchased the product.
Connect the FP-4135 On-Board Volumetric Flow Detector to the Ono Sokki DF-2200 On-Board Flow Meter/FM Series
Digital Flowmeter with the flow signal cable (FP-0015/0016/0017) and the temperature signal cable (FP-0025/0026/0027)
following the procedure below.
[1]
Connecting flow detecting part and signal processing part of the FP-4135 On-Board Volumetric Flow Detector
Connect the cable for the flow detecting part of FP-4135 On-Board Volumetric Flow Detector to the SENSOR
connector of the signal processing part.
[2]
Connecting signal processing part and flow detecting part of FP-4135 On-Board Volumetric Flow Detector
Connect the flow signal cable to the FLOW connector in the signal processing part of FP-4135 On-Board
Volumetric Flow Detector.
[3]
Connecting temperature sensor of FP-4135 On-Board Volumetric Flow Detector to temperature signal cable
Connect the temperature signal cable to the temperature sensor connector of the FP-4135 On-Board Volumetric
Flow Detector.
[4]
Connecting flow signal cable and temperature signal cable to Ono Sokki DF-2200 On-Board Flow Meter/FM Series
Digital Flowmeter
First, turn off the Ono Sokki DF-2200 On-Board Flow Meter or the FM Series Digital Flowmeter.
Then, connect the flow signal cable to the Ono Sokki DF-2200 On-Board Flow Meter or the FM Series Digital
Flowmeter rear panel FLOW connector.
In the same manner, connect the temperature signal cable to the TEMP connector.
!
CAUTION
•
The combination of the flow rate detecting part and the signal processing part of the FP-4135 On-Board Volumetric Flow
Detector is pre-defined. Connect after verifying that the manufacturing number of the flow rate detecting part and the signal
processing part matches.
•
Do not put any unnecessary force on the cables or connectors when wiring.
5.3 Configuring the FACTOR
The FACTOR value of the FP-4135 On-Board Volumetric Flow Detector is “10000”. This value must be entered
(configured) into the Ono Sokki DF-2200 On-Board Flow Meter or FM Series Digital Flowmeter before measurement.
For details on configuration, refer to the attached instruction manual for the Ono Sokki DF-2200 On-Board Flow Meter
or FM Series Digital Flowmeter.
Checking FACTOR value
The FACTOR value is imprinted on the nameplate of the FP-4135 On-Board Volumetric Flow Detector.
FACTOR value