61
When you see this:
You can do this:
Hypoglycemic
threshold
Modify the
Hypoglycemic threshold
by entering the desired
number. Please note whether the Hypoglycemic threshold is
displayed in mg/dL (or mmol/L) units and as a Plasma (or Whole
Blood) value.
The Hypoglycemic threshold must be set below the lower limits of
all Before-Meal and After-Meal target ranges.
Depending on locale one of the following default hypoglycemic
threshold settings is provided:
■
67 mg/dL (3.7 mmol/L) when using plasma values, and 60mg/dL
(3.3 mmol/L) when using whole blood values
or
■
70 mg/dL (3.9 mmol/L) when using plasma values, and
63 mg/dL (3.5 mmol/L) when using whole blood values.
To change the Hypoglycemic threshold, enter the desired threshold
value in the field. The Hypoglycemic threshold must be from:
■
50 to 90 mg/dL (2.8 to 5 mmol/L) when using plasma values
or
■
45 to 80 mg/dL (2.5 to 4.4 mmol/L) when using whole blood
values.
Notes:
■
The Hypoglycemic threshold will change whenever you shift
between Whole Blood and Plasma display options. You may
switch display options in the General Tab of the Preferences
screen. OneTouch
®
DMS will convert the value automatically.
■
You can customize the hypoglycemic value for individual
patients in the Patient Setup General screen.