PROGRAMMING
Cycle Time
The standard and recommended default disinfection cycle is 5 minutes. However, the
cycle time can be customized to better meet facility protocols.
Note: The amount of
killing or inactivation UV-C radiation does is proportional to the amount of exposure time.
TO SET CYCLE TIME:
•
From the Home screen, choose the Settings button (see Settings Screen
Components).
•
Choose the Set Cycle button.
Note: The time is set by arrowing up or down. The format
is automatically programmed in MM:SS.
•
Once desired lamp life cycle has been chosen, select the check mark to set.
PROGRAMMING ITEMS TO NOTE:
•
Controller cannot be programmed if EMERGENCY STOP buttons are pressed in. Pressing
EMERGENCY STOP button(s) disconnects all power, even to the controller.
•
If disconnected from power, the BioShift
®
chamber does not have to be reprogrammed.
Sleep Timer
In order to assure the longest life, the BioShift
®
chamber will automatically enter sleep
mode after 30 minutes of inactivity. This setting can be changed by following this path:
•
From the Home screen, choose the Settings button (see Settings Screen
Components).
•
Choose the Sleep Time option and set to desired time.
•
Select the check mark to set.