7
EN
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules
may result in damage, inefficient nebulization or infection. For the instructions, refer to
“Cleaning and disinfecting” section.
• Clean and disinfect the nebulizer kit, mask or mouthpiece before using them when as
follows:
- the first time after purchase.
- if the device has not been used for a long period of time
- if more than one person uses the same device
• Be sure to wash or wipe the parts after use, and ensure that they are thoroughly
disinfected and dried, and stored in a clean location.
• Do not leave the cleaning solution to dry on the parts.
Caution:
Indicates a potentially hazardous situation which if not avoided, may result in minor or
moderate injury, or physical damage.
(Usage)
• In order to prevent strangulation by the power cord or air tube, or injuries caused by
mishandling, please provide close supervision when this device is used by, on, or near
children or invalids.
• Make sure that the parts are attached correctly.
• Make sure that the vaporizer head is correctly installed before use.
• Make sure that the air filter is correctly attached.
• Make sure that the air filter is clean. If the air filter has been used for more than 60 days,
replace it with a new one.
• Do not spill liquid or medication on the compressor.
• Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions or shake
it while in use.
• Do not use or store the device while the air tube is creased.
• Use only original nebulizing parts, air tube, air filter and air filter cover.
• Do not add more than 7 ml of medication to the medication tank.
• Do not carry or leave the nebulizer kit while the medication tank contains medication.
• Do not leave the device unattended with infants or persons who cannot express their
consent.
• Do not subject the device or the parts to any strong shocks such as dropping on the floor.
• Do not distort the vaporizer head, or jab the nozzle of the medication tank with a pin or any
sharp object.
• Do not insert fingers or objects inside the compressor.
• Do not disassemble or attempt to repair the compressor or AC adapter.
• Do not leave the device or its parts where it will be exposed to extreme temperatures or
changes in humidity, such as in a vehicle during the summer, or where it will be exposed to
direct sunlight.
• Do not block the air filter cover.
• Do not use the device while sleeping or if drowsy.
• Do not block the slit between the cap and the inhalation air inlet.
• Approved for human use only.
• If the device is used continuously, the service life of the device may be shortened.
• Limit use to 20 minutes at a time, and allow a 40 minute interval before using the device again.
• When using the device, the main unit may become hot.
NE-C801S-KDRU.book Page 7 Monday, August 27, 2018 4:15 PM
Summary of Contents for NE-C24
Page 27: ...27 EN NE C801S KDRU book Page 27 Monday August 27 2018 4 15 PM...
Page 30: ...44 46 1 40 2 1 2 3 41 NE C801S KDRU book Page 30 Monday August 27 2018 4 15 PM...
Page 31: ...3 4 41 5 42 39 43 NE C801S KDRU book Page 31 Monday August 27 2018 4 15 PM...
Page 34: ...34 o OMRON 53 NE C801S KDRU book Page 34 Monday August 27 2018 4 15 PM...
Page 35: ...35 RU 60 ee 45 7 20 40 NE C801S KDRU book Page 35 Monday August 27 2018 4 15 PM...
Page 36: ...36 30 OMRON HEALTHCARE 40 C 55 NE C801S KDRU book Page 36 Monday August 27 2018 4 15 PM...
Page 37: ...37 RU 1 NE C801S KDRU book Page 37 Monday August 27 2018 4 15 PM...
Page 39: ...39 RU 2 4 44 46 7 48 2 1 1 2 3 NE C801S KDRU book Page 39 Monday August 27 2018 4 15 PM...
Page 40: ...40 2 2 2 1 2 1 2 3 2 7 4 5 1 2 NE C801S KDRU book Page 40 Monday August 27 2018 4 15 PM...
Page 41: ...41 2 RU 6 1 2 3 7 1 2 NE C801S KDRU book Page 41 Monday August 27 2018 4 15 PM...
Page 42: ...42 3 1 45 2 20 40 3 45 45 NE C801S KDRU book Page 42 Monday August 27 2018 4 15 PM...
Page 43: ...43 3 RU 4 1 2 3 NE C801S KDRU book Page 43 Monday August 27 2018 4 15 PM...
Page 44: ...44 4 4 1 1 2 3 2 40 41 4 5 6 7 NE C801S KDRU book Page 44 Monday August 27 2018 4 15 PM...
Page 45: ...45 4 RU 4 2 4 1 46 A 1 2 3 B 15 20 NE C801S KDRU book Page 45 Monday August 27 2018 4 15 PM...
Page 47: ...47 RU 5 1 2 3 6 NE C801S KDRU book Page 47 Monday August 27 2018 4 15 PM...
Page 48: ...48 7 60 1 2 3 OMRON OMRON NE C801S KDRU book Page 48 Monday August 27 2018 4 15 PM...
Page 49: ...49 RU 8 30 Omron 45 20 20 40 NE C801S KDRU book Page 49 Monday August 27 2018 4 15 PM...
Page 57: ...57 RU NE C801S KDRU book Page 57 Monday August 27 2018 4 15 PM...
Page 58: ...58 NE C801S KDRU book Page 58 Monday August 27 2018 4 15 PM...
Page 59: ...59 RU NE C801S KDRU book Page 59 Monday August 27 2018 4 15 PM...