8. Troubleshooting
In case of any of the below problems occur during use, first check that no other electrical
device is within 30 cm. If the problem persists, please refer to the table below.
Note:
If the suggested remedy does not solve the problem, do not try to repair the device - no parts of
the unit are user serviceable.
Return the unit to an authorized OMRON retail outlet or distributor.
Problem
Cause
Remedy
Nothing happens when the
power switch is pressed.
Is the AC adapter connected
correctly to the an electrical
outlet and the compressor?
Check that the plug is
inserted in an electrical
outlet. Unplug then reinsert
the plug if necessary.
No nebulization or low
nebulization rate, when the
power is on.
Is there medication in the
medication tank?
Add the correct amount of
medication to the medication
tank.
Is there too much/little
medication in the medication
tank?
Is the vaporizer head missing
and not assembled correctly?
Attach the vaporizer head
correctly.
Is the nebulizer kit
assembled correctly?
Assemble the nebulizer kit
correctly.
Is the nozzle blocked?
Make sure that the nozzle is
free of blockages.
Is the nebulizer kit tilted at a
sharp angle?
Make sure that the nebulizer
kit is not tilted at an angle of
more than 45 degrees.
Is the air tube connected
correctly?
Make sure that the air tube is
correctly connected to the
compressor and nebulizer kit.
Is the air tube folded or
damaged?
Make sure that the air tube
does not contain kinks.
Is the air tube blocked?
Make sure that the air tube is
free of blockages.
Is the air filter dirty?
Replace the air filter with a
new one.
The device is abnormally
loud.
Is the air filter cover attached
correctly?
Attach the air filter cover
correctly.
The device is very hot.
Is the compressor covered?
Do not cover the compressor
with any type of cover during
use.
Is the device operating
continuously over
20 minutes?
Limit use to 20 minutes at a
time, and allow a 40 minute
interval before using the
device again.
NE-C801S-KDRU.book Page 21 Monday, August 27, 2018 4:15 PM
Summary of Contents for NE-C24
Page 27: ...27 EN NE C801S KDRU book Page 27 Monday August 27 2018 4 15 PM...
Page 30: ...44 46 1 40 2 1 2 3 41 NE C801S KDRU book Page 30 Monday August 27 2018 4 15 PM...
Page 31: ...3 4 41 5 42 39 43 NE C801S KDRU book Page 31 Monday August 27 2018 4 15 PM...
Page 34: ...34 o OMRON 53 NE C801S KDRU book Page 34 Monday August 27 2018 4 15 PM...
Page 35: ...35 RU 60 ee 45 7 20 40 NE C801S KDRU book Page 35 Monday August 27 2018 4 15 PM...
Page 36: ...36 30 OMRON HEALTHCARE 40 C 55 NE C801S KDRU book Page 36 Monday August 27 2018 4 15 PM...
Page 37: ...37 RU 1 NE C801S KDRU book Page 37 Monday August 27 2018 4 15 PM...
Page 39: ...39 RU 2 4 44 46 7 48 2 1 1 2 3 NE C801S KDRU book Page 39 Monday August 27 2018 4 15 PM...
Page 40: ...40 2 2 2 1 2 1 2 3 2 7 4 5 1 2 NE C801S KDRU book Page 40 Monday August 27 2018 4 15 PM...
Page 41: ...41 2 RU 6 1 2 3 7 1 2 NE C801S KDRU book Page 41 Monday August 27 2018 4 15 PM...
Page 42: ...42 3 1 45 2 20 40 3 45 45 NE C801S KDRU book Page 42 Monday August 27 2018 4 15 PM...
Page 43: ...43 3 RU 4 1 2 3 NE C801S KDRU book Page 43 Monday August 27 2018 4 15 PM...
Page 44: ...44 4 4 1 1 2 3 2 40 41 4 5 6 7 NE C801S KDRU book Page 44 Monday August 27 2018 4 15 PM...
Page 45: ...45 4 RU 4 2 4 1 46 A 1 2 3 B 15 20 NE C801S KDRU book Page 45 Monday August 27 2018 4 15 PM...
Page 47: ...47 RU 5 1 2 3 6 NE C801S KDRU book Page 47 Monday August 27 2018 4 15 PM...
Page 48: ...48 7 60 1 2 3 OMRON OMRON NE C801S KDRU book Page 48 Monday August 27 2018 4 15 PM...
Page 49: ...49 RU 8 30 Omron 45 20 20 40 NE C801S KDRU book Page 49 Monday August 27 2018 4 15 PM...
Page 57: ...57 RU NE C801S KDRU book Page 57 Monday August 27 2018 4 15 PM...
Page 58: ...58 NE C801S KDRU book Page 58 Monday August 27 2018 4 15 PM...
Page 59: ...59 RU NE C801S KDRU book Page 59 Monday August 27 2018 4 15 PM...