5.Handling Errors and Problems
5. Handling Errors and Problems
5.1
Error Messages
Error Display
Cause
Remedy
Movement during
measurement
Repeat measurement. Remain still
and do not talk during
measurement.
Refer to section 3.3.
Air plug disconnected.
Insert the air plug securely.
Refer to section 3.2.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.2.
Clothing is interfering with
the arm cuff.
Remove any clothing interfering
with the arm cuff.
Refer to section 3.2.
Air is leaking from the arm
cuff.
Replace cuff with new one.
Refer to Chapter 7.
The arm cuff was inflated
above 299 mmHg when
inflating the cuff manually.
Remove the arm cuff and take
another measurement.
Refer to section 3.3.
Battery power is low.
Replace all four “AAA” batteries
with new ones.
Refer to section 2.1.
Blinks or appears
continuously
Care and maintenance