7

How to Assemble

A. Clean and sterilize all unit parts before the first use or after a long storage period.

B. Fill the medication cup before plugging in the unit and turning on the power.

UNIT ASSEMBLY

2. Attach the 

medication cup
to the back of the
unit.  Push the
medication cup
vertically until a
click is heard.

How to Fill Medication Cup

Be sure that all parts are clean before filling 
medication cup. Place the prescribed medication
directly into the medication cup. Never use more
than 5 ml of solution.

3. The mesh cap is attached as follows:

A. Place the cap position indicator onto the unit 

turning counterclockwise to 10 o’clock

B. Then turn the cap clockwise  to 12 o’clock until a 

click is heard to assure proper attachment.

4. Attached the mask or mouthpiece directly over the

mesh cap unit a click is heard. Remember to handle 
the mesh cap with care to prevent breakage.

5. Attach the AC adapter plug to the "AC Adapter Jack"

for use with electrical current.The unit can also be 
battery powered by using 4 "AA" batteries as
described on page 13.

1. Remove the

cover by gently
pushing in an
upward direction.

Summary of Contents for MicroAir NE-U03V

Page 1: ...INSTRUCTION MANUAL R PORTABLE NEBULIZER MODEL NE U03V...

Page 2: ...ne into an aerosol form This change is done using vibration The Micro Air was designed with your needs in mind and weighs less than one pound The Micro Air runs with a standard AC adapter or with 4 AA...

Page 3: ...on or near children or compromised individuals 3 Use this product only for its intended use as described in this manual Do not use attachments not recommended by the manufacturer 4 Never operate this...

Page 4: ...IT The shape of the plug may vary from country to country AC Adapter Unit Cover Mask Medication Cup Mesh Cap Extra Cap Included MicroAir Unit Mouthpiece AC Jack Vibration Horn On Off Button Rinse Bott...

Page 5: ...page 10 3 WARNING NEVER submerge or wet the AC power unit or cord An electrical shock may result 6 WARNING NEVER operate while driving a car or boat 7 WARNING Do not force the bottom of the battery co...

Page 6: ...ater to prevent damage 11 CAUTION Assure that the proper adapter is used for connection to any electrical outlet to prevent damage to the unit or injury 12 CAUTION Check the medication cup and mouthpi...

Page 7: ...e medication cup Never use more than 5 ml of solution 3 The mesh cap is attached as follows A Place the cap position indicator onto the unit turning counterclockwise to 10 o clock B Then turn the cap...

Page 8: ...is controlled by holding the button down for four 4 seconds Press the ON OFF button again to turn off the unit NOTES A The green operation light will be on during continuous operation B A blinking gr...

Page 9: ...igh pitched sound when all medication has been nebulized Quickly turn the unit off when this sound is heard it shows that there is no more medication in the unit Pour a mixture of one part vinegar and...

Page 10: ...ilable disinfectant C Parts may also be boiled for 10 minutes D All parts must be rinsed thoroughly and air dried Clean the main unit casing using a cloth moistened with water and a few drops of non a...

Page 11: ...l of the mist stops Turn the unit OFF and replace the unit cover 1 Follow the steps listed above 2 Soak the mesh cap in water vinegar solution for 10 to 15 minutes after removing from the unit 3 Swish...

Page 12: ...tion cup to the unit as described in the operating instructions Unit Cover Attach cover to the unit by sliding the cover down until it clicks into place Mouthpiece or Mask The mouthpiece or mask can b...

Page 13: ...atteries will need to be replaced if either of these conditions occur A When the green operation light blinks during continuous use B The unit shuts off while being battery operated Replace all four 4...

Page 14: ...l outlet Is the AC adapter connected to the unit Are the batteries inserted correctly Are the batteries charged Are alkaline batteries inserted Is there medication in the medication cup See above for...

Page 15: ...MODEL U03 4 Plastic Mouthpiece AC Adapter MODEL U03 5 Battery Lid MODEL U03 9 AC Adapter 120v 60 Hz Unit Cover MODEL U03 8 Carrying Case MODEL U03 10 Pediatric Mask MODEL U03 11 Rinse Bottle MODEL U10...

Page 16: ...uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that...

Page 17: ...ery small holes is it likely to clog during use A The mesh cap will operate without clogging with proper cleaning after each use as described in this manual Q What should I do if my mesh clogs during...

Page 18: ...e mouthpiece Is this normal A Yes Up to 02 ml can remain in the mouthpiece from condensation that results from the aerosol process Q Should I take my treatment with the MicroAir set on continuous flow...

Page 19: ...correct defects in material or workmanship covered by this limited warranty THIS LIMITED WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER ORAL WRITTEN EXPRESS OR IMPLIED THERE ARE NO WARRANTIES OF...

Page 20: ...HEALTHCARE INC 300 Lakeview Parkway Vernon Hills Illinois 60061 Phone 1 800 634 4350 www omronhealthcare com Copyright OMRON HEALTHCARE INC NE U03 INST Rev B OMRON uses recycled paper to help preserve...

Reviews: