MESH CLEANING
NOTE:
The mesh cap is the most delicate part of the MicroAir. It is important
not to directly touch the mesh cap. Follow the cleaning instructions to avoid
damaging the mesh cap and keep it in good condition.
1. Remove the medication cup and rinse with water after turning the unit OFF.
2. Nebulize all medication remaining in the mesh cap and transducer after
removing the medication cup. Turn the unit off when the aerosol stops.
3. Nebulize a mixture of one part vinegar and three parts distilled water for
45 to 60 seconds through the medication cup.
4. Remove the medication cup and leave the unit ON. Keep the unit running
until all of the mist stops. Turn the unit OFF and replace the unit cover.
1. Follow the steps listed above.
2. Soak the mesh cap in water/vinegar solution for 10 to 15 minutes after
removing from the unit.
3. Swish the meshcap in a blow of warm water.
DO NOT rinse mesh cap
directly under a faucet as this will cause damage.
4. Replace mesh cap back onto nebulizer and turn the unit on until the
aerosol mist stops. Quickly turn the unit OFF. Replace the unit cover.
1. Follow the steps listed above in section A.
2. Place mesh cap in a microwavable cup with one cup of distilled water.
3. Microwave on high for 3 minutes
4. Soak mesh cap in warm water for 30 minutes
5. Replace mesh cap back on nebulizer and run until the aerosol mist stops.
Turn the unit OFF. Replace the unit cover.
A. After Every Cleaning
B. Daily Cleaning
C. Weekly Cleaning For Difficult Medications