K3MA-J

Parameter Initialization

This function returns all of the parameters to their initial values.

Use this to reset the K3MA-J after returning it to its factory-set condi-
tion.

Average Processing

Average processing stabilizes displayed values to minimize flicker by
averaging the fluctuating input signals. Average processing can be
performed for the measurement values in either of four steps (OFF, 2
times, 4 times, or 8 times).

This is useful for ignoring rapid fluctuations, e.g., eliminating spike
noise.

Hysteresis (Comparative Output 
Models Only)

The hysteresis of comparative outputs can be set to prevent chatter-
ing in the output when the measurement value fluctuates finely near
the OUT value.

Zero-limit Function

The zero-limit function changes any value below the set value to
zero. This is useful when you want to change negative values to zero
rather than display them, or when you want to make the display in the
smallest part of the input range zero.

Changing the Display Color

The color of the value displayed can be set to either red or green. For
comparative output models (

K3MA-J-A2)

, the display color can be

set to change from green to red, or from red to green, according to
the status of the comparison criterion.

Display Auto-return Time

This function automatically returns the display to the operation level’s
current value if no keys are pressed for a preset time (called the dis-
play auto-return time).

Move-to-Protect-Level Time

The time required to shift to the protect level can be set as desired.

Forced-zero Function

It is possible to shift from a value to the zero point with one touch of
the Up Key on the front panel (for example, when adjusting reference
values).

Note:

Used only for releasing the forced-zero with the Protect menu.

MAX/MIN Display

The maximum and minimum measurement (display) values from the
time the power is turned ON until the current time can be stored and
displayed. This is useful, for example, when measuring the maximum
value.

Parameter

Setting value

Meaning

init

off

---

on

Initializes all parame-
ters.

Parameter

Setting value

Meaning

=-lim

off

OFF: No zero-limit

on

ON: Zero-limit

lim-p

0

 to 

99

0 to 99: Zero-limit value

Upper limit (hi

g

h actin

g

)

Set 

v

al

u

e

Meas

u

rement 

v

al

u

e

Hysteresis

O

u

tp

u

t

O

N

OFF

Display

Zero-limit 
setting 

v

al

u

e

Inp

u

t

8.781

100.05

113.36

OUT1 

v

al

u

e

OUT2 

v

al

u

e

Green

Red

Red

UP

MAX/MIN

LEVEL

MODE

SHIFT

C

u

rrent 

v

al

u

e

MAX 

v

al

u

e

MI

N

 

v

al

u

e

MAX/MI

N

MAX/MI

N

MAX/MI

N

Max

Min

MAX/MIN

LEVEL

MODE

SHIFT

UP

MAX/MIN

LEVEL

MODE

SHIFT

UP

MAX/MIN

LEVEL

MODE

SHIFT

UP

Summary of Contents for K3MA-J 100-240VAC

Page 1: ...CE marking Model Number Structure Model Number Legend 1 Input Type J DC voltage current 2 Comparative Output Model None No output A2 2 relay contact outputs SPST NO 3 Supply Voltage 100 240VAC 100 to...

Page 2: ...on resistance Vibration 10 to 55 Hz Acceleration 50 m s2 5 min each in X Y and Z directions for 10 sweeps Shock resistance 150 m s2 100 m s2 for relay contact outputs 3 times each on 3 axes 6 directio...

Page 3: ...red OUT type change upper limit lower limit upper lower limit Average processing simple average Output Relays 2 SPST NO Delay in comparative outputs 750 ms max Degree of protection Front panel NEMA4X...

Page 4: ...ge input For current input OUT1 OUT2 COM COM Current input Terminal No Name Description Operation power Connects the operation power supply Analog input Connects the voltage or current analog input Ou...

Page 5: ...ng value Meaning Parameter Setting value Meaning inp 1 19999 to 99999 Input value for dsp 1 dp Display four digits after decimal point dsp 1 19999 to 99999 Input value for inp 1 Display three digits a...

Page 6: ...set value The three types of output operations shown above can be combined as desired The following are examples of possible combinations T lights to indicate that the parameter is valid for teaching...

Page 7: ...or from red to green according to the status of the comparison criterion Display Auto return Time This function automatically returns the display to the operation level s current value if no keys are...

Page 8: ...ge the level 6 Mode Key Used to allow the main indicator to indicate parameters sequentially 7 Shift Key Used to enable a set value to be changed When changing a set value this key is used to move alo...

Page 9: ...packing is provided with the Controller Wiring Precautions Use crimp terminals Tighten the terminal screws to a torque of approximately 0 5 N m To avoid the influence of noise route signal lines and...

Page 10: ...e the forced zero The Z indicator will turn OFF If forced zero protection is turned ON in the protect level the Up Key cannot be used to set a forced zero Forced zero protection is ON in the default s...

Page 11: ...etting level 1234 5 Power ON Operation level Protect level Initial setting level Advanced function setting level Time set by user See note 1 s min 1 s min 1 s min Flashing stops if key is released Con...

Page 12: ...UT2 upper limit value OUT2 lower limit value OUT2 value Some parameters are not displayed for certain models The K3MA J will stop measurement if the level is changed to the initial setting level or ad...

Page 13: ...d function setting level Enter password 0169 Parameter initialization Average processing OUT1 hysteresis OUT2 hysteresis Models with the compar ative output function MODE MODE MODE MODE MODE MODE MODE...

Page 14: ...tting change lockout Forced zero shift lockout Parameter Setting Operation level Current value display Set value display oapt 0 Allowed Allowed 1 Allowed Allowed 2 Allowed Prohibited Parameter Setting...

Page 15: ...Repair is necessary Consult your OMRON sales representative 5 e111 EEPROM memory error When this error is displayed press the Level Key for 3 sec onds and the settings will be restored to the factory...

Page 16: ...ther in no event shall liability of Omron Companies exceed the individual price of the Product on which liability is asserted Suitability of Use Omron Companies shall not be responsible for conformity...

Page 17: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Omron K3MA J A2 24VAC VDC K3MA J 24VAC VDC...

Reviews: