13
3.1 Taking a Measurement
6
Apply the sound sensor to the patient’s skin.
Place and adhere the sound sensor firmly.
Not adhering
Place and adhere firmly
7
The device will beep twice and begin measuring.
When the sound sensor is correctly placed on the body, the
device gives twice the sound "beep" and the measurement
starts automatically.
After the sound sensor is applied to the skin, the sensor will
flash white while taking the measurement.
Beep, Beep
Note
• Ensure that the patient DOES NOT talk, cry or move while the measurement is being
taken.
• Ensure that the patient continues to breathe normally.
8
Wait until the measurement is completed.
When the measurement is completed, the device beeps 3 times.
The maximum measurement time is 30 seconds.
Note
KEEP the sensor in close contact and DO NOT move the device until the device beeps 3
times. Otherwise the result may be inaccurate.