Troubleshootin
g
In case any of the below problems occur durin
g
use, first check that no other electrical unit
is within 30 cm. If the problem persists, please refer to the table below.
Trouble
What to inspect
How to correct
The
u
nit inflates to
a
b
normally high (or
lo
w
) press
u
re.
Is the c
u
ff
w
rapped correctly?
Wrap the c
u
ff correctly, and
meas
u
re again.
(Refer to Page 11 and 12.)
Did the patient mo
v
ed his/her arm
or
b
ody d
u
ring inflation?
Ask the patient not to mo
v
e his/
her arm or
b
ody d
u
ring
meas
u
rement, and meas
u
re
again.
Does this tro
ub
le occ
u
r to a
specific patient?
For some patients
w
ith
arrhythmia, this monitor may not
inflate properly.
Set the P-SET at 30 to 40 mmHg
higher than the systolic press
u
re
of the patient, then meas
u
re.
The monitor cannot
meas
u
re
b
lood
press
u
re.
Meas
u
red
v
al
u
es are
a
b
normally high (or
lo
w
).
First of all, check the patient's
condition.
After checking the patient
w
ith the
stethoscope as needed, take
appropriate meas
u
re as in the
"List of error codes".
(Refer to Page 24.)
Did the patient mo
v
ed his/ her
arm or
b
ody d
u
ring
meas
u
rement?
Ask the patient not to mo
v
e his/
her arm or
b
ody d
u
ring
meas
u
rement, and meas
u
re
again.
Does the patient ha
v
e
arrhythmia?
Check the patient
w
ith a
stethoscope.
Is the size of the c
u
ff correct and
is it
w
rapped correctly?
Select the c
u
ff according to the
patient's arm circ
u
mference,
w
rap
it correctly, then meas
u
re again.
(Refer to Pages 11 and 12.)
Is the le
v
el of the
b
rachi
u
m to
w
hich the c
u
ff is
w
rapped at the
same le
v
el as the heart?
Keep the le
v
el of the
b
rachi
u
m to
w
hich the c
u
ff is
w
rapped at the
same le
v
el as the heart, then
meas
u
re again.
(Refer to Page 12.)
How to Measure Blood Pressure
HEM-907-C3_D_M.fm Page 28 Wednesday, July 4, 2018 12:11 PM