Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, K5 was used on 85 subjects for determination
of diastolic blood pressure.
• This monitor is clinically investigated according to the requirements of
ISO 81060-2:2013 and complies with ISO 81060-2:2013 and ISO 81060-
2:2018+A1:2020.
• This device has not been validated for use on pregnant patients.
• IP classification is degrees of protection provided by enclosures in
accordance with IEC 60529. The device and AC adapter are protected
against solid foreign objects of 12 mm diameter and greater such as a finger.
The AC adapters (HEM-ADPTW5/HHP-CM01) are protected against
vertically falling water drops which may cause issues during a normal
operation. Other AC adapters (HHP-OH01 and HHP-BH01) are protected
against oblique falling water drops which may cause issues during a normal
operation.
Dimensions
Monitor : Approximately 107 (w) mm × 79 (h) mm × 141 (l) mm
(4 1/4" × 3 1/8" × 5 1/2")
Arm cuff : Approximately 145 mm × 594 mm (air tube: 750 mm)
(5 3/4" × 23 1/2" (air tube: 29 1/2"))
Cuff circumference
22 to 42 cm (9" to 17")
Contents
Monitor, arm cuff, battery set, instruction manual, quick start guide
Applied part
Type BF (cuff)
Protection against
electric shock
Internally powered ME equipment (When using only the batteries)
Class II ME equipment (Optional AC adapter)