PM-1794-01-06/2018 

Operation Mode  

Continuous operation

MDD Classification 

Class II a

Packaging contents 

Main unit, Arm cuff (M size), AC Adapter, Instruction 
Manual, Guarantee card

 
Power supply 

AC adapter 

Input voltage range: AC 100 V to 240 V 
Frequency: 50/60 Hz 
Output voltage range: DC 6 V ±0.5 V 
Rated Output Current: 0.7 A

Dry cell battery 

Type: AA batteries, x4 
Approx. 250 measurements 
• Measurement conditions 
- New batteries (AA high-performance manganese) 
- Ambient temperature of 23°C (73.4°F) 
- Using M-size cuff 
- SYS120 / DIA80 / PR60 
- One 5-minute cycle consisting of “cuff 
measurement time + wait time”

 

Environmental Conditions

 

Operating Conditions 

Temperature range: 5 to 40°C (41 to 104°F) 
Humidity range: 15 to 85%RH (not condensed) 
Atmospheric pressure: 700 to 1060hPa 

Storage and transportation 

Temperature range: -20 to 60°C (-4 to 140°F) 
Humidity range: 10 to 95%RH (not condensed) 
Atmospheric pressure: 500 to 1060hPa 

 

Non-Invasive Blood Pressure (NIBP)

 

Measurement technology 

Oscillometric

Measurement method 

Dynamic Linear Deflation method 

Pressure display range 

0 to 300 mmHg

Pressure display accuracy 

Within ±3 mmHg

NIBP measurement range 

SYS 60 to 250 mmHg 
DIA 40 to 200 mmHg 
PULSE 40 to 200 /min

NIBP accuracy* 

Maximum mean error within ±5 mmHg 
Maximum standard deviation within 8 mmHg 

Pulse rate accuracy

Within ±5 % of reading

Reference Standard 

EN1060-1:1995+A2:2009 
EN1060-3:1997+A2:2009 
EN80601-2-30:2010+A1:2015 
EN ISO 81060-2:2013

* Comparison with auscultation method performed by a trained professional. 
   DIA determined by the auscultation method is “K5”.

 

 
 

Summary of Contents for HBP-1120

Page 1: ...1 0 23 55 44 701 www omron healthcare com BANKERS BANK OF TOKYO MITUSBISHI HOLLAND N V ACC NR 063 56 15 320 IBAN NL95BOTK0635615320 SWIFT BOTKNL2X CHAMBER OF COMMERCE NR 342 10306 BTW VAT NR NL8132 86...

Page 2: ...el code HBP 1120 HBP 1120 E LEGAL MANUFACTURER OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko KYOTO 617 0002 Japan PRODUCION FACILITY OMRON DALIAN Co Ltd No 3 Song Jiang Road Economic and Techni...

Page 3: ...HXA GCFXL PBE AC Adapter HHP CM01 HHP CM01 E AC Adapter UK HHP BFH01 HHP BFH01 E General description OMRON s HBP 1120 HBP 1120 E blood pressure monitor is a professional medical device meant for meas...

Page 4: ...for use This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population Measurement Parameter Non Invasive Blood Pressure Pulse rate Features...

Page 5: ...a bare arm or over thin clothing 3 Select the mode of application Normal Mode Auscultation Mode 4 Press start 5 Note the blood pressure and pulse rate in the patients documents Review the Instruction...

Page 6: ...104 F Humidity range 15 to 85 RH not condensed Atmospheric pressure 700 to 1060hPa Storage and transportation Temperature range 20 to 60 C 4 to 140 F Humidity range 10 to 95 RH not condensed Atmosphe...

Page 7: ...rdboard Instruction Manual Paper Materials used for accessories Cuff Item Material Outside cloth Nylon Polyester Inner cloth Polyester Air bag Flexible urethane sheet Velcro tape hook Nylon Velcro tap...

Page 8: ...8 PM 1794 01 06 2018 Title page of Instruction Manual...

Page 9: ...9 PM 1794 01 06 2018 Unfolded package design of main device and accessories Main device Information in white square on package artwork...

Page 10: ...10 PM 1794 01 06 2018 AC adapter HHP CM01 AC Adapter with UK plug HHP BFH01...

Page 11: ...EN 1060 3 1997 A2 2009 EN 60601 1 2006 A1 2013 EN 60601 1 6 2010 A1 2015 EN 80601 2 30 2010 A1 2015 EN 62304 2006 A1 2015 EN 62366 2008 A1 2015 EN ISO 10993 1 2009 AC 2010 EN ISO 10993 5 2009 EN ISO...

Page 12: ...oxygen therapy device or in an environment where combustible gas may be generated Do not use in combination with magnetic resonance imaging MRI equipment If MRI is to be performed remove cuff connect...

Page 13: ...he measurement results of the unit when judging the patient s condition Only trained healthcare providers should use this device Do not allow patients to operate this device Properly connect the conne...

Page 14: ...en cleaning the unit turn off the power and disconnect the AC adapter from the unit After cleaning the unit make sure it is completely dry before connecting to a power outlet Do not spray pour or spil...

Page 15: ...on wrap the cuff as indicated in the cautionary points in this manual Do not connect the NIBP cuff or cuff joint to a luer lock adapter Do not bend cuff tube during inflation and deflation particularl...

Page 16: ...easurement Disposal As there is a risk of environmental pollution follow your applicable national and local legal regulations regarding disposal or recycling of this equipment The main constituents of...

Page 17: ...utor Servicing The HBP 1120 must be maintained to ensure functionality and to secure the safety of patients and operators Daily checks and maintenance should be performed by the operator In addition q...