Operation and care
Cleaning and disinfecting should be performed in accordance with your facility’s
infection control practice.
Cleaning of the device:
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%,
or ethyl alcohol (disinfection alcohol) diluted to 80 v/v% or less and wrung out.
Do not wipe the Power connector or allow it to become wet.
Warnings
Read all the information in the instruction manual and any other literature included in
the box before using the device.
Storage / Setup
Warnings
• Install the unit in a location close to a power outlet where the AC adapter can be
easily disconnected. If the power cannot be quickly disconnected when an abnormal
condition occurs, an accident or fire may result.
• Do not use the cuff or AC adapter to lift the unit, it can also cause the unit to
malfunction.
• If the unit has broken down, contact contact your OMRON retail dealer or
distributor.
• Do not use in combination with a hyperbaric oxygen therapy device, or in an
environment where combustible gas may be generated.
• Do not use in combination with magnetic resonance imaging (MRI) equipment. If
MRI is to be performed, remove cuff connected to the unit from the patient.
• Do not use with a defibrillator.
• Do not install the unit in the following locations:
- Locations subject to vibration such as ambulances and emergency helicopters.
- A location where there is gas or flame.
- A location where there is water or steam.
- A location where chemicals are stored.
• Do not use at extremely high temperature, high humidity, or high altitude. Use only
within the required ambient conditions.
• Do not subject the unit to intense shock.
• Do not place heavy objects on the AC adapter cable, or allow the unit to sit on the
cord.
• Clinical testing has not been conducted on newborn infants and pregnant women.
Do not use on newborn infants and pregnant women.
• Do not plug in or unplug the AC adapter with wet hands.
• During measurement, make sure that no mobile phone or any other electrical
devices that emit electromagnetic fields is within 30 cm of this monitor. This may
result in incorrect operation of the monitor and/or cause an inaccurate reading.