Electrostatic
discharge (ESD)
IEC 61000-4-2
<
<
<
<
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Electric fast
transient/burst
IEC 61000-4-4
`<
power supply
lines
"<
input/output lines
`<
power supply
lines
"<
input/output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
"<
differential mode
`<
common mode
"<
differential mode
`<
common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
for 0.5 cycle
<5 % U
T
for 0.5 cycle
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
HBP-1100 requires
continued operation
during power mains
interruptions, it is
recommended that the
HBP-1100 be powered
from an uninterruptible
power supply or
batteries.
40 % U
T
for 5 cycles
40 % U
T
for 5 cycles
70 % U
T
for 25 cycles
70 % U
T
for 25 cycles
U
T
?&
of test unit
<5 % U
T
for 5 sec.
<5 % U
T
for 5 sec.
Power frequency
G#^#[J
magnetic field
IEC 61000-4-8
3 A/m (r.m.s)
3 A/m (r.m.s)
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
% ]
T
is the a.c. mains voltage prior to application of the test level.