3.
Using the Unit
3.1
Taking a Reading
Notes:
• Make sure the probe cover is attached correctly.
• It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the
highest temperature.
Ear Measurement Mode
1.
Press the ON/MEM button.
All symbols appear on the display.
Then the display shown at the right will appear with 2 beeps.
2.
Insert the probe into the ear as far as it comfortably goes in
the direction of the eardrum.
Notes:
• Gently pull the ear back to straighten the ear canal and position the probe into the ear so it is
snug, aiming towards the membrane of the eardrum to obtain an accurate reading.
• Holding the unit too long may cause a higher ambient temperature reading of the probe. This
could make the body temperature measurement lower than usual.
MC-521-E_A_M_main.book Page 7 Tuesday, March 26, 2013 2:41 PM
Summary of Contents for Gentle Temp 521
Page 24: ...6 Some Useful Information 23 EN MC 521 E_A_M_main book Page 23 Tuesday March 26 2013 2 41 PM ...
Page 48: ...6 Informations utiles 47 FR MC 521 E_A_M_main book Page 47 Tuesday March 26 2013 2 41 PM ...
Page 72: ...6 Nützliche Hinweise 71 DE MC 521 E_A_M_main book Page 71 Tuesday March 26 2013 2 41 PM ...
Page 96: ...6 Informazioni utili 95 IT MC 521 E_A_M_main book Page 95 Tuesday March 26 2013 2 41 PM ...
Page 120: ...6 Información útil 119 ES MC 521 E_A_M_main book Page 119 Tuesday March 26 2013 2 41 PM ...
Page 144: ...6 Nuttige informatie 143 NL MC 521 E_A_M_main book Page 143 Tuesday March 26 2013 2 41 PM ...
Page 168: ...6 Полезная информация 167 RU MC 521 E_A_M_main book Page 167 Tuesday March 26 2013 2 41 PM ...
Page 215: ......
Page 216: ...HEM 6200 E_ en_TOC fm Page 14 Tuesday May 26 2009 2 00 PM ...