background image

E4PA-N

 

Ultrasonic Displacement Sensor

 

5

Operation

Setup and Operating Procedure

The E4PA-N is factory-set for the decrease mode for analog output characteristics and for the maximum sensing range. These settings enable
scaling with the connected device (e.g., the K3NX). In this case, only step 4 is required. 

To change the analog output range, follow steps 1 to 4 below.

1. Preparation for Setup

Note:

Do not insert the Setting Plug yet. It must be inserted within
five minutes after the power is turned ON, however, or the set-
up will not be possible. If that happens, power up again and
start over.

2. Setting Sensing Range

3. Setting Analog Output Characteristics 
Mode

4. Setting Temperature Compensation Mode

Note:

If the reflection from the Sensor is extremely weak, or if the sensing object is set outside of the sensing range, the error indicator (red) will
flash to show that the sensing range is not accurately set. If the Setting Plug is left removed for five minutes or longer, temperature com-
pensation will be disabled (status before setting). 

If it becomes necessary to change the settings for step 2 (Sensing Range) or step 3 (Analog Output Characteristics Mode) after steps 1 through
4 have been completed, the contents of all the previous settings will be remembered, so only the items to be changed need to be reset. Go
through the steps in order, from 1 to 4, and reset only the items to be changed.

Step

Setting/Operation

1

Remove the Setting Plug.

2

Turn ON the power. 
(See note.)

3

Proceed to Step 2 

Setting Sensing Range.

Step

Setting/Operation

1

Place the sensing object at the nearest point within the sensing 
range.

2

Insert the Setting Plug at the A1 
position. (Align the notch on the 
Sensor with the A1 triangular mark 
on the Setting Plug.)

The power indicator (green) and 
the A1 RUN indicator (yellow) with 
flash, showing that the near dis-
tance has been correctly set.

3

Remove the Setting Plug. (In this 
state, the setting is confirmed and 
recorded.)

4

Place the sensing object at the farthest point within the sensing 
range.

5

Insert the Setting Plug at the A2 
position. (Align the notch on the 
Sensor with the A2 triangular mark 
on the Setting Plug.)

The power indicator (green) and 
the A2 RUN indicator (yellow) will 
flash, showing that the far distance 
has been correctly set.

6

Remove the Setting Plug. (In this 
state, the setting is confirmed and 
recorded.)

7

Proceed to Step 3 

Setting Analog Output Characteristics Mode.

50 mm (for E4PA-LS50-M1-N)

Sensing object

A1

A2

E3

E2

A1 position

500 mm (for E4PA-LS50-M1-N)

Sensing object

A2 position

A1

A2

E3

E2

Step

Setting/Operation

1

Insert the Setting Plug at the E2/E3 
position. (Align the notch on the Sen-
sor with the E2/E3 triangular mark on 
the Setting Plug.)

Note:

Check the input impedance of connected devices
and the analog output in 

Ratings/Characteristics 

(see

page 2

).

2

After the analog output characteristics have been set, remove the 
Setting Plug.

3

Proceed to Step 4 

Setting Temperature Compensation Mode.

Step

Setting/Operation

1

Insert the Setting Plug at the T posi-
tion. (Align the notch on the Sensor 
with the T triangular mark on the Set-
ting Plug.)

2

Leave the Setting Plug inserted where it is. (This completes the set-
up.)

Note:

To maintain the temperature compensation, leave
the Setting Plug inserted during operation. If it is re-
moved from the E4PA, the temperature compensa-
tion will be disabled.

E2/E3 position

A1

A2

E3

E2

A1

A2

E3

E2

A1

A2

E3

E2

 A2 RUN indicator (yellow) flashing:

 Analog output characteristics: Decrease mode 

In this mode, the analog output decreases 
(20 to 4 mA/10 to 0 V) as the  
sensing object nears the E4PA 
within the sensing range.

  A1 RUN indicator (yellow) flashing:

 Analog output characteristics: Increase mode

 A1 and A2 RUN indicators (yellow) flashing at the same time:

 Analog output characteristics: Output from 0 mm

In this mode, the analog output 
increases (4 to 20 mA/0 to 10 V) 
as the sensing object nears the 
E4PA within the sensing range.

In this mode, the analog output 
decreases (20 to 4 mA/10 to 0 V) 
as the sensing object nears the 
E4PA between the farthest 
sensing point and 0 mm.

The mode changes when the Setting Plug  
is reinserted at the E2/E3 position. 

The mode changes when the Setting Plug  
is reinserted at the E2/E3 position. 

T position

Downloaded from

Arrow.com.

Downloaded from

Arrow.com.

Downloaded from

Arrow.com.

Downloaded from

Arrow.com.

Downloaded from

Arrow.com.

Summary of Contents for E4PA-LS200-M1-N

Page 1: ...cu racy of 1 mm or better has been achieved A setting plug with a temperature compensation function makes it easy to set the sensing range The sensing range can be set by the Setting Plug which is inc...

Page 2: ...180 kHz Approx 85 kHz Approx 65 kHz Response time 63 ms max 195 ms max 440 ms max 850 ms max Power supply voltage 10 to 30 VDC ripple p p 10 max Power consumption 1 800 mW max Analog output Current o...

Page 3: ...ance Y mm Distance X m Y X 1 000 800 600 400 200 0 200 400 600 800 1 000 1 2 3 4 5 6 7 Distance Y mm Distance X m 1 4 2 3 5 E4PA N 0 V Ri 500 Rv 1 000 3 1 2 4 5 10 to 30 V 4 to 20 mA 0 to 10 V Mutual...

Page 4: ...ctor for Sensor I O 5 poles Notch on Sensor RUN indicator yellow Power indicator green RUN indicator yellow Error indicator red Position Function A1 Nearest sensing point A1 A2 Farthest sensing point...

Page 5: ...notch on the Sensor with the A2 triangular mark on the Setting Plug The power indicator green and the A2 RUN indicator yellow will flash showing that the far distance has been correctly set 6 Remove...

Page 6: ...etec tion capability of the Sensor will vary depending on the surface con ditions of the object If the surface irregularities are no more than 0 2 mm then the specified sensing distance will apply Mat...

Page 7: ...indicator red 36 See note 2 25 22 25 5 12 128 27 5 40 dia M30 1 5 M12 1 Two clamping nuts Indicator See note Setting Plug Note 1 RUN indicator yellow Power indicator green Error indicator red 2 This...

Page 8: ...t to change without notice ALL DIMENSIONS SHOWN ARE IN MILLIMETERS To convert millimeters into inches multiply by 0 03937 To convert grams into ounces multiply by 0 03527 D099 E1 OMRON Corporation Ind...

Page 9: ...not described in this catalog Nuclear energy control systems combustion systems railroad systems aviation systems medical equipment amusement machines vehicles safety equipment and installations subj...

Reviews: