ERROR MESSAGES AND TROUBLESHOOTING
Error Display
Cause
Solution
Movement during
measurement and
the arm cuff has not
been inflated
sufficiently.
Repeat measurement. Remain still
and do not talk during measurement.
Refer to “Taking a Measurement”.
If “E2” appears repeatedly, inflate
the arm cuff manually until it is
30 to 40 mmHg above your previous
measurement result.
Refer to “Taking a Measurement”.
The arm cuff was
inflated exceeding the
maximum allowable
pressure, and then
deflated automatically
when inflating the arm
cuff manually.
Do not touch the arm cuff and/or bend
the air tube while taking a
measurement. Do not inflate the arm
cuff more than necessary.
Refer to “Taking a Measurement”.
Movement during
measurement.
Repeat measurement. Remain still
and do not talk during measurement.
Refer to “Taking a Measurement”.
Clothing is interfering
with the arm cuff.
Remove any clothing interfering with
the arm cuff.
Refer to “Applying the Arm Cuff”.
Device error.
Contact Customer Service.