GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP710CANN is intended for use in the electromagnetic environment specified below. The
customer or the user of this OMRON BP710CANN should assure that it is used in such environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The OMRON BP710CANN uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The OMRON BP710CANN is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not Applicable.
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
Not Applicable.