GUIDANCE AND MANUFACTURER’S DECLARATION

(Table 1)

Guidance and manufacturer’s declaration – electromagnetic emissions

Emissions test

Harmonic emissions 

IEC 61000-3-2

Voltage fluctuations/

flicker emissions

IEC61000-3-3

(Table 2)

Guidance and manufacturer’s declaration – electromagnetic immunity

Immunity test IEC 60601

Test level

Compliance

level

Electromagnetic environment 

– guidance

Immunity test IEC 60601

Test level

Compliance

level

Electromagnetic environment 

– guidance

(Table 6)

Recommended separation distance between portable and mobile RF 

communications equipment and the OMRON BPM 

Separation distance according to frequency of 

transmitter in meter

Not Applicable.
Not Applicable.
Not Applicable.
Not Applicable.
Not Applicable.

0.12
0.38

1.2
3.8

12

0.23
0.73

2.3
7.3

23

0.01

0.1

1

10

100

Output Power of

Transmitter in 

Watt

(Table 4)

Guidance and manufacturer’s declaration – electromagnetic immunity

OMRON Wrist Blood Pressure Monitor (BPM)

Information for accompanying documents in the scope of IEC60601-1-2:2007

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile 

(cellular) telephones, medical devices in use may be susceptible to 

electromagnetic interference from other devices. Electromagnetic interference 

may result in incorrect operation of the medical device and create a potentially 

unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibil-

ity) with the aim to prevent unsafe product situations, the IEC60601-1-2 

standard has been implemented. This standard defines the levels of immunity 

to electromagnetic interferences as well as maximum levels of electromagnetic 

emissions for medical devices.
Medical devices manufactured by OMRON Healthcare conform to this 

IEC60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:

 The use of accessories and cables other than those specified by OMRON,

 with the exception of cables sold by OMRON as replacement parts for

 internal components, may result in increased emission or decreased  

 

  immunity of the device.

 The medical devices should not be used adjacent to or stacked with other

 equipment.

 In case adjacent or stacked use is necessary, the medical device should be 

 observed to verify normal operation in the configuration in which it will be

 used.

• Refer

 to further guidance below regarding the EMC environment in which    

  the device should be used.

 The MEDICAL ELECTRICAL E

Q

UIPMENT BPM needs special  

 

 precautions regarding EMC and needs to be installed and put into service    

  according to the EMC information provided in this documentations.

 The Essential Performance of the BPM is to measure a blood pressure and a  

  pulse rate and using the memory function. 
The BPM may be interfered with by other equipment, even if that other 

equipment complies with CISPR EMISSION requirements.

OMRON BPM is intended for use in the electromagnetic environment 

specified below. The customer or the user of this OMRON BPM should assure 

that it is used in such environment.

Electromagnetic environment 

– guidance

Group 1

Class B

RF emissions CISPR 11

RF emissions CISPR 11

The OMRON BPM uses 

RF energy only for its internal 

function. Therefore, its RF emissions 

are very low and are not likely to cause 

any interference in nearby electronic 

equipment.

The OMRON BPM 

is suitable for use in all establishments, 

including domestic establishments and 

those directly connected to the public 

low-voltage power supply network that 

supplies buildings used for domestic 

purposes.

OMRON BPM is intended for use in the electromagnetic environment 

specified below. The customer or the user of this OMRON BPM should assure 

that it is used in such environment.

OMRON BPM is intended for use in the electromagnetic environment 

specified below. The customer or the user of this OMRON BPM should assure 

that it is used in such environment.

Compliance

Electrostatic

discharge (ESD)

IEC 61000-4-2

Electrical fast

transient/burst

IEC 61000-4-4

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

Surge 

IEC 61000-4-5

Voltage dips, 

short 

interruptions 

and voltage 

variations on 

power supply 

inputlines 

IEC 61000-4-11

Power 

frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

Floor should be wood, concrete, 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be at 

least 30 %.

Not Applicable.

Not Applicable.

Power frequency magnetic fields 

should be at levels characteristic 

of a typical location in a typical 

commercial or hospital 

environment.

Not Applicable.

Portable and mobile RF 

communications

equipment should be used no 

closer to any part of

the OMRON BPM 

including cables, than the 

recommended separation 

distance calculated from the 

equation appropriate to the 

frequency of the transmitter.

±6 kV contact

±8 kV air

±2 kV for 

power

supply lines

±1 kV for

input/output

lines

Not

Applicable.

±1 kV

line(s) to 

line(s)

±2 kV

line(s) to 

earth

Not

Applicable.

<5 % 

U

T

(>95 % dip 

in 

U

T

)

for 0.5 cycle

40 % 

U

T

(60 % dip in 

U

T

)

for 5 cycles

70 % 

U

T

(30 % dip in 

U

T

)

for 25 cycles

Not

Applicable.

<5 % 

U

T

(>

95 % dip 

in 

U

T

)

for 5 sec.

±6 kV contact

±8 kV air

3 A/m

3 A/m

Note: 

U

T

 is the A.C. mains voltage prior to application of the test level.

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2:  These guidelines may not apply in all situations. Electromagnetic

 

propagation is affected by absorption and reflection from structures,

 

objects, and people.

a

  Field strengths from fixed transmitters, such as base stations for radio (cellular/  

  cordless) telephones and land mobile radio, AM and FM radio broadcast, and  

  TV broadcast cannot be predicted theoretically with accuracy. To assess the   

  electromagnetic environment due to fixed RF transmitters, an electromagnetic  

  site survey should be considered. If the measured field strength in the location  

  in which the OMRON BPM is used exceeds the applicable 

  RF compliance level above, the OMRON BPM should be observed to  

 

  verify normal operation. If abnormal performance is observed, additional    

  measures may be necessary, such as reorienting or relocating the OMRON    

 BPM.

b

  Over the frequency range 150 kHz to 80 MHz, field strengths should be less  

  than 3 V/m.

3 V rms

150 kHz to

80 MHz

3 V/m

80 MHz to

2.5 GHz

3 V/m

Not 

Applicable.

Recommend 

separation distance

Not Applicable.

OMRON BPM is intended for use in an electromagnetic environment in which 

radiated RF disturbances are controlled. The customer or the user of this 

OMRON BPM can help prevent electromagnetic interference by maintaining a 

minimum distance between portable and mobile RF communications 

equipment (transmitters) and the OMRON BPM as recommended below, 

according to the maximum output power of the communications equipment.

150 kHz to 80 MHz

Not Applicable.

80 MHz to 800 MHz

d

 = 1.2 √P

800 MHz to 2.5GHz

d

 = 2.3 √P

For transmitters rated at a maximum output power not listed above, the 

recommended separation distance 

d

 in meters (m) can be estimated using the 

equation applicable to the frequency of the transmitter, where 

P

 is the maximum 

output power rating of the transmitter in watts (W) according to the transmitter 

manufacturer.

Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency  

 

range applies

Note: These guidelines may not apply in all situations. Electromagnetic  

 

 

propagation is affected by absorption and reflection from structures,  

 

 

objects, and people.

Not

Applicable.

Not

Applicable.

d

 = 1.2 √P 

 

80 MHz to 800 MHz

d

 = 2.3 √P 

 

800 MHz to 2.5 GHz

where P is the maximum output 

power rating of the transmitter in 

watts (W) according to the

transmitter manufacturer and 

d

 is 

the recommended separation 

distance in meters (m).

Field strengths from fixed RF 

transmitters as determined by an 

electromagnetic site survey,

a

should be less than the compli-

ance level in each frequency 

range.

b

Interference may occur in the 

vicinity of equipment marked 

with the following symbol:

TROUBLESHOOTING TIPS

PROBLEM

No power.

No display appears on the unit.

Readings appear too high or too 

low.

CAUSES AND SOLUTIONS

Replace both batteries with new ones at the 

same time. Check the battery installation for 

proper placement of the battery polarities.

Review the section “Battery Installation”.
Blood pressure varies constantly. Many factors 

including stress, time of day, how you wrap the 

cuff, where you position your wrist, may affect 

your blood pressure. 

Review the sections “Before Taking a 

Measurement” and “Taking a Measurement”.

LIMITED WARRANTY

Your BP654 Wrist Blood Pressure Monitor, excluding the batteries, is warranted to 

be free from defects in materials and workmanship appearing within 5 years from 

the date of purchase, when used in accordance with the instructions provided with 

the monitor. The above warranty extends only to the original retail purchaser.
We will, at our option, replace without charge any monitor covered by the above 

warranty. Replacement is our only responsibility and your only remedy under the 

above warranty.
To obtain warranty service contact Customer Service by calling 

1-800-634-4350

 

for the address of the inspection center and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone 

number, and description of the specific problem. Pack the product carefully to 

prevent damage in transit. Because of possible loss in transit, we recommend 

insuring the product with return receipt requested.

THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN 

CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY 

DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, 

INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND 

FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND 

OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE 

LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS 

WARRANTY.

OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER 

SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, 

EXPENSES OR DAMAGES.

This warranty provides you with specific legal rights, and you may have other 

rights which vary from state to state. Some states do not allow limitations on how 

long an implied warranty lasts, so the above limitation may not apply to you. Some 

states do not allow the exclusion or limitation of incidental or consequential 

damages, so the above limitation or exclusion may not apply to you.

FOR CUSTOMER SERVICE

Visit our web site at:                                                     www.omronhealthcare.com /

 www.omronhealthcare.ca

Call toll free:                                                                                     1-800-634-4350

SPECIFICATIONS

Model:
Display:
Measurement Range:

Accuracy:

Inflation:
Deflation:
Measurement Method:
IP Classification :
Power Source:
Battery Life:
Operating Temperature 
/Humidity:
Storage Temperature 
/Humidity/Air Pressure:
Main Unit Weight:

Main Unit Dimensions:

Measurable 
Circumference of Wrist:
Memory:
Contents:

Protection against 
Electric Shock:
Applied Part:

BP654         HEM-6320T-Z
LCD Digital Display

Pressure: 0 mmHg to 299 mmHg 

Pulse: 40 beats/min to 180 beats/min
Pressure: ±3 mmHg or 2% of reading

Pulse: ±5% of reading
Automatic by electric pump
Automatic rapid deflation
Oscillometric method
IP 22
2 “AAA” alkaline batteries 1.5V
Approximately 300 uses with 2 new alkaline batteries

50ºF to 104ºF (10ºC to 40ºC) /15% RH to 90% RH

-4ºF to 140ºF (-20ºC to 60ºC) /10% RH to 95% RH

700 hPa to 1060 hPa
Approximately 3 1/2 oz. (100 g)

not including batteries
Approximately 3 1/2''(w) × 2 3/8''(l) × 1''(d)

(89 mm × 61 mm × 25 mm) (not including the wrist cuff)

Approximately 5 1/4" to 8 1/2" (13.5 cm to 21.5 cm)

Up to 100 readings
Main Unit, Storage Case, 2 “AAA” Alkaline Batteries, 

Instruction Manual, and Quick Start Guide

Internally powered ME equipment

Manufactured for :

OMRON HEALTHCARE Co., Ltd

53, Kunotsubo, Terado-cho

Muko, Kyoto, 617-0002 JAPAN

Distributed by :

OMRON HEALTHCARE, INC.

1925 West Field Court

Lake Forest, IL 60045 U.S.A.

www.omronhealthcare.com

© 2014 OMRON HEALTHCARE, INC.

Made in Vietnam

FCC CAUTION

Changes or modifications not expressly approved by the party responsible for 

compliance could void the user's authority to operate the equipment.

Note:

This equipment has been tested and found to comply with the limits for a Class B 

digital device, pursuant to part 15 of the FCC Rules. These limits are designed to 

provide reasonable protection against harmful interference in a residential 

installation. This equipment generates, uses and can radiate radio frequency 

energy and, if not installed and used in accordance with the instructions, may 

cause harmful interference to radio communications. However, there is no 

guarantee that interference will not occur in a particular installation. If this 

equipment does cause harmful interference to radio or television reception, which 

can be determined by turning the equipment off and on, the user is encouraged to 

try to correct the interference by one or more of the following measures:

• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the 

receiver is connected.

• Consult the dealer or an experienced radio/TV technician for help.

This transmitter must not be co-located or operated in conjunction with any other 

antenna or transmitter.

This equipment complies with FCC/IC radiation exposure limits set forth for an 

uncontrolled environment and meets the FCC radio frequency (RF) Exposure 

Guidelines and RSS 102 of the IC radio frequency (RF) Exposure rules. This 

equipment has very low levels of RF energy that are deemed to comply without 

testing of specific absorption ratio (SAR).

This device complies with Part 15 of FCC Rules and Industry Canada 

license-exempt RSS standard(s). Operation is subject to the following two 

conditions: (1) this device may not cause interference, and (2) this device must 

accept any interference, including interference that may cause undesired operation 

of this device.

This Product operates in the unlicensed ISM band at 

2.4GHz. In case this Product is used around the other 

wireless devices including microwave and wireless LAN, 

which operate same frequency band of this Product, there 

is a possibility that interference occurs between this 

Product and such other devices. If such interference 

occurs, please stop the operation of other devices or 

relocate this Product before using this Product or do not 

use this Product around the other wireless devices.

The 

Bluetooth

®

 Smart

 word mark and logos are registered 

trademarks owned by Bluetooth SIG, Inc. and any use of 

such marks by OMRON HEALTHCARE Co.,Ltd. is under 

license. Other trademarks and trade names are those of 

their respective owners.

Apple, the Apple logo, iPad, and iPhone are trademarks of Apple Inc., registered 

in the U.S. and other countries. 

App Store is a service mark of Apple Inc.

Android and Google Play are both trademarks of Google Inc.

Samsung and Galaxy S are both registered trademarks of Samsung Electronics 

Co., Ltd.

FCC/IC STATEMENT AND TRADEMARKS

ERROR MESSAGES

SYMBOL

CAUSE

CORRECTION

Wrist cuff not applied 

correctly.

Apply the wrist cuff correctly. Refer to

“Applying the Wrist Cuff ”.

Recommend to replace the batteries 

with new ones ahead of time.

Refer to “Battery Installation”.

Do not hold the wrist cuff. Remain still

and do not talk during measurement.

Refer to “Taking a Measurement”.

Remain still until the measurement is

complete. Refer to “Taking a

Measurement”.

Contact Customer Service.

Apply the wrist cuff correctly. Refer to

“Applying the Wrist Cuff ”.

Remain still and do not talk during

measurement. Refer to “Taking a

Measurement”.

Movement during 

measurement.

Wrist cuff not applied 

correctly or movement 

during measurement.

Arm position changed 

during measurement.

Device error.

CARE AND MAINTENANCE

To keep your digital blood pressure monitor in the best condition and protect the 

unit from damage, follow the directions listed below:

CAUTION

• Do not forcefully bend the cuff. Do not fold tightly.

• Clean the monitor with a soft and dry cloth.

• Do not use any abrasive or volatile cleaners.

• Wipe clean on the surface of the cuff with a soft, moistened cloth with a 

dilution of neutral detergent.

• Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the 

inside well.

CAUTION

Do not wash the device and any components or immerse them in water. Do 

not subject the monitor to extreme hot or cold temperatures, humidity or 

direct sunlight.

CAUTION

Store the device and the components in a clean, safe location.

CAUTION

Do not subject the monitor to strong shocks, such as dropping the unit on the floor.

CAUTION

Remove the batteries if the unit will not be used for three months or longer. 

Always replace all the batteries with new ones at the same time.

CAUTION

Use only OMRON authorized parts and accessories. Parts and accessories 

not approved for use with the device may damage the unit.

CAUTION

Changes or modification not approved by the manufacturer will void the user 

warranty. Do not disassemble or attempt to repair the unit or components.

CAUTION

Do not store the device in the following situations:

• If the device is wet.

• Locations exposed to extreme temperatures, humidity, direct sunlight, dust 

or corrosive vapors such as bleach.

• Locations exposed to vibrations, shocks or where it will be at an angle.

Data cannot be sent.

The destination device is too far away from the 

monitor. After checking that there are no sources 

of interference nearby, move the device to a 

distance within 16 ft. (5 m) of the monitor.
The 

Bluetooth

 function on the destination 

device is turned off. Turn on the 

Bluetooth

 

function and try sending the data again.

Pairing (registering) has not been completed. 

Perform pairing (registration). Refer to “To Pair 

This Device With a Smartphone”.
The application on the destination device is not 

ready. Check the application then try sending the 

data again. Refer to “To Pair This Device With a 

Smartphone”. If the Err symbol still lights after 

checking the application, contact customer service.

NOTES:

 

• 

These specifications are subject to change without notice.

 

• 

In the clinical validation study, the 5th phase was used on 85 subjects 

for determination of diastolic blood pressure.

 

• 

This device has not been validated for use in pregnancy.

• IP classification is degrees of protection provided by IEC 60529.

Batteries are low.

Replace the two batteries.

Refer to “Battery Installation”.

Batteries are depleted.

The 

Bluetooth

 function on the monitor is not 

turned on. Press the      /       button, turn on the 

Bluetooth

 function and try sending the data again.

Movement during 

measurement.

Carefully read and repeat the steps in 

section “Taking a Measurement”.

Irregular heartbeats are 

detected.

Please remove the wrist cuff, and wait 

2-3 minutes. Reapply the wrist cuff and 

take another measurement.

Repeat the steps in section “Taking a 

Measurement”.

If this error continues to appear, contact 

your physician.

Summary of Contents for BP654

Page 1: ...d If there are only 2 readings in the memory for that period the average will be based on these 2 readings If there is 1 reading in the memory for that period this is displayed as the average 1 Push down the hook of the battery cover and pull downward 2 Install 2 AAA size batteries so the positive and negative polarities match the polarities of the battery compartment as indicated 3 Replace the ba...

Page 2: ...and description of the specific problem Pack the product carefully to prevent damage in transit Because of possible loss in transit we recommend insuring the product with return receipt requested THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT AND OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND ...

Page 3: ...uede firmemente cerrada COLOCACIÓN DE LAS PILAS COLOCACIÓNDELBRAZALETEPARALAMUÑECA Pantalla Compartimiento de las pilas Botón de memoria Botón START STOP Botón de transferencia botón de configuración de fecha hora Brazalete de muñeca Indicador de nivel de presión arterial BARRA Pantalla de fecha hora Presión arterial sistólica Símbolo de lectura promedio Símbolo de posición Presión arterial diastó...

Page 4: ...uras Unidad principal estuche 2 pilas alcalinas AAA manual de instrucciones y guía de inicio rápido Equipo ME de alimentación interna Fabricado para OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPAN Distribuido por OMRON HEALTHCARE INC 1925 West Field Court Lake Forest IL 60045 U S A www omronhealthcare com 2014 OMRON HEALTHCARE INC Fabricado en Vietnam PRECAUCIÓN SEGÚN LA ...

Page 5: ...émoire Symbole de dégonflement Symbole de transfert Symbole de mémoire pleine Symbole OK Indicateur de positionnement Symbole de la mémoire Symbole d erreur de mouvement Symbole de rythme cardiaque irrégulier INTRODUCTION Merci d avoir acheté le tensiomètre de poignet BP654 d OMRONMD Ce tensiomètre compact bénéficie d une grande facilité d utilisation à la maison au travail ou en déplacement Idéal...

Page 6: ...ES GARANTIES DE QUALITÉ MARCHANDE ET D ADAPTATION À UN USAGE PARTICULIER LES GARANTIES TACITES ET LES AUTRES MODALITÉS POUVANT ÊTRE IMPOSÉES PAR LA LOI SI DE TELLES GARANTIES OU MODALITÉS EXISTENT SONT LIMITÉES DANS LEUR DURÉE À LA PÉRIODE DE GARANTIE EXPRESSE MENTIONNÉE AUX PRÉSENTES OMRON NE POURRA ÊTRE TENU RESPONSABLE NI DES PERTES DE JOUISSANCE NI D AUTRES COÛTS DÉPENSES OU DOMMAGES PARTICULI...

Reviews: