LVS-7500 External System Operations Manual
LVS-7500 External System Operations Manual
Page 104 of 129
o
UsePassFailMethod=1 – Displays the Alarm Matrix in the “Pass/Fail View” which is I/O based on the
pass or fail status of the entire inspection.
[Alarm matrix]
Line01=Good read ,I/O line 1,100ms,
Line02=Stop motion ,I/O line 4,100ms,
Line03=Grade warning, I/O line 2,100ms,
Line04=Background ,I/O line 3,100ms,do not stop
Line05=Bypass mode ,I/O line 3,100ms,do not stop
Line06=Database engine, I/O line 3,100ms,immediately
Line07=Die cut ,I/O line 3,100ms,do not stop
Line08=Duplicate ,I/O line 3,100ms,do not stop
Line09=Foreground ,I/O line 3,100ms,do not stop
Line10=Gap ,I/O line 3,100ms,do not stop
Line11=Matrix ,I/O line 3,100ms,do not stop
Line12=Mismatch ,I/O line 3,100ms,do not stop
Line13=Not assessed ,I/O line 3,100ms,do not stop
Line14=Not synced ,I/O line 3,100ms,do not stop
Line15=No read ,I/O line 3,100ms,do not stop
Line16=Quality ,I/O line 3,100ms,do not stop
Line17=Range ,I/O line 3,100ms,do not stop
Line18=Sequence ,I/O line 3,100ms,do not stop
Line19=Wrong length ,I/O line 3,100ms,do not stop
UseAcknowledgementIO=1
UsePassFailMethod=0
Aperture Settings:
The aperture settings are used to calculate barcode grading. These should not be changed by a customer
unless requested by a
n
Omron Microscan
representative.
[Aperture]
ApertureScaling=1.0
IgnoreGS1Rules=0
SharpenMatrixSize=5
SharpenValue=0.5
Blemish Settings:
The Blemish Settings are user configurable.
•
ColorCode, Debug (default is 0), FloodDiff, and FloodSize are only to be used when requested by a
n
Omron Microscan
representative.
•
MaxErrorsBeforeQuit – When a given sector reaches more than the number of errors defined in this field,
the LVS-7500 stops reporting the errors. For example, If MaxErrosBeforeQuit=20, the LVS-7500 stops
reporting errors after a given sector reaches 20 errors.
NOTE:
This value is configurable; however, the
value must limited to keep from causing processing overflow.
•
MaxThumbnailsPerSector limits how many thumbnails are used for a single sector’s errors.
•
ScrollMax controls the maximum size of a blemish. Value is in inches.
•
ShowDieCutCrossHatch is used by
Omron Microscan
technicians. Default is 0
•
ShowMatrixDieCutControls is used to make the sensitivity controls for Matrix and Die Cut available during
setup.
•
SquareSize is the size of boxes that any given Blemish sector will be cropped into.