© 2010 Omnicell, Inc.
OmniRx Service Guide/
67-2023 Rev D
Electronics Sled
1-21
Standardization Changes
Standardization Changes
To standardize the peripherals that Omnicell products use, the Seiko printer and PS/2 keyboards
are being replaced by APS printers and USB keyboards with a numeric pad. The
is
swapped out in this procedure.
1. Lift and prop-up the console if it is not already open.
2. Disconnect all console cables.
3. Remove the console.
with the new, wider one using three
.
Figure 1-37.
Cable cover bracket
5. Place new console into position, then prop it open.
6. Re-connect the cables. Most will go back where the old cables were with a couple of exceptions.
goes into J33 instead of J32.
goes to the USB port (vs. the PS/2 connection).
on JP1 from pins 1-2 to 3-4.
Final Steps
After the service is performed, return the OmniRx to production status.
1. Lift and prop-up the console if it is not already open.
2. Re-connect the battery cable.
3. Close and lock the console.
4. Plug in the power cable, then restart the OmniRx.