Omni Mixer User Manual Download Page 9

In the event that the motor drive connect and/

or the flexible rubber coupling needs to be ser-

viced or replaced, the aluminum coupling can 

be removed using the following procedure:

1.  Unplug motor unit from 

the wall.

2.  Insert flat edge of 

combination wrench 

(included in motor 

tool kit) into the 

slots of the alumi-

num coupling (see 

figure to right).

3.  Using wrench for 

leverage, loosen 

aluminum cou-

pling by twisting it 

counter-clockwise.

Installation of the coupling is the reverse of 

these steps.

Maintenance

Motor Drive Unit

Removal of

Aluminum Coupling

The motor drive housing can be cleaned period-

ically with a mild detergent. Never use solvents 

to clean the unit or accessories. While resistant 

to most laboratory chemicals, care should be tak-

en to wipe off any concentrated and potentially 

damaging liquids.

Once per month the aluminum coupling adapter 

(see figure to right) of the homogenizer should 

be removed and cleaned. After cleaning, a good 

lubricant should be applied to the threads.

To a large measure, how well your homogenizer 

performs its job is determined by the condition 

of the carbon contact brushes (P/N 61810). If 

brushes require replacement, the unit may run 

at a reduced speed and make unusual noises, 

or it may run sporadically and lose power. If this 

occurs with your Omni Mixer or Omni Macro, 

unplug the cord, loosen the brush caps one at 

a time and check each brush. If the brushes are 

less than 1/8” long and the ends of the brushes 

(which contact the commutator) are rough and/

or pitted, the brushes no longer make adequate 

contact with the commutator and should be re-

placed (check both brushes since both brushes 

generally do not wear at the same rate). If one 

brush is worn replace both brushes. Make sure 

that the brushes are installed properly. The 

brushes have a curved surface at the end which 

must match the curvature of the commutator. 

The curvature of the commutator is the same as 

the curvature of the motor housing.

If the brushes are not installed properly they will 

not contact the commutator and the motor will 

not run. After replacing the brushes, make sure 

to remove the flexible rubber coupling, place the 

motor on a clean surface, and run it freely (no-

load, without a generator or chamber assembly 

attached) for 15 to 30 minutes before using 

the unit to process samples. This will allow the 

brushes to “seat” properly and will extend the 

life of each set of brushes. This will also extend 

the total life of your motor since the commutator 

surface will wear evenly.

Motor Brush Maintenance

Motors are equipped with sealed ball bearings. 

Under normal use they require no additional lu-

brication. A tube of lubricant (P/N 61971) is in-

cluded with every drive unit and should be used 

to lubricate the threads of all connecting thread-

ed parts, as well as the drive pin at the tip of the 

floating shaft .Lubricant should not be used to 

lubricate moving parts or bearings.

Bearings and Lubrication

Grounding Instructions

In the event of a malfunction or breakdown, 

grounding provides a path of least resistance 

for electric current to reduce the risk of electric 

shock. This instrument is equipped with an elec-

tric cord which is grounded to the chassis hous-

ing. The plug must be plugged in to an outlet 

that is properly installed and grounded in accor-

dance with all local codes and ordinances.

Summary of Contents for Mixer

Page 1: ...Omni Mixer Macro User Manual OMNI MIXER HOMOGENIZER www omni inc com...

Page 2: ...SAVE THIS OWNER S MANUAL The Omni Mixer and Omni Macro has been engineered for economical functionality as well as safety however basic safety precautions and common sense must always be demonstrated...

Page 3: ...Instruction Manual 17258 17258 1 Overview 1 2 3 4 5 1 Support Post 2 Coupling Adapter 3 Support Base 4 Clamping Knob Assembly 5 Motor Support Housing Description P N QTY Gen Probe Multi Tool 00 402 2...

Page 4: ...robes can be used in open containers or in sealed chamber assemblies Sample process ing efficiency is effected by Amount of material processed vs size and speed of the generator probe Container geomet...

Page 5: ...ot thread fully into the generator probe adapter then unscrew the generator probe from the generator probe adapter Slightly rotate the rotor shaft collar on the generator probe Then return to step 2 4...

Page 6: ...eleased Determining Motor Speed To determine motor speed during operation a tachometer not in cluded may be inserted into the Omni Mixer or Omni Macro through the top of the motor as shown by removing...

Page 7: ...g wear or part failure NOTE Inspect glass or plastic chambers for any scratches or cracks and do not use if any are found Eye protection and utilization of a secondary container is recommended to prev...

Page 8: ...n the assembly securely 2 Attach the chamber assembly to the motor by screw ing the chamber cover into the coupling adapter 3 If the chamber assembly does not thread fully into the coupling adapter th...

Page 9: ...commutator and should be re placed check both brushes since both brushes generally do not wear at the same rate If one brush is worn replace both brushes Make sure that the brushes are installed prop...

Page 10: ...d turned on but is not functioning Check power cord connectors Check wall socket for power to the outlet Motor is turned ON and makes a buzz ing sound but is not working Brushes may need to be replace...