5.3
Connection
To connect the
SAA2
system to the main device:
1. Make sure that the main device used with the
SAA2
is switched off.
2. Press the
Emergency Stop
button on the main device or test system (in case of
VOTANO 100
the
Emergency button
on
SAB1
safety box) and lock the main device with a physical key (
CPC 80
and
CPC 100
) or by pressing the software lock button (
TESTRANO 600
and
VOTANO 100
).
NOTE!
CIBANO 500
and
COMPANO 100
do not have a separate physical key nor do they have a
software lock function.
WARNING
Death or severe injury caused by high voltage or current possible.
Always place the
SAA2
control unit in the working area.
3. Place the
SAA2
control unit close to the main device and position the
SAA2
signal lamps so that
they are clearly visible for all operators and if possible also so that the lamps surround the test area.
Use the magnetized surface to place the signal lamp on metallic objects, or fix it by using the tripod
screw, or by using the mounting ring. Make sure that the signal lamps are firmly in place and that the
Emergency Stop
buttons are easily accessible.
Figure 11: Mounting options
4. Connect the delivered power supply to the
SAA2
control unit and connect the mains plug of the
power supply to the power outlet
.
WARNING
Death or severe injury caused by high voltage or current possible.
Use SAA2 only with the original power supply delivered by OMICRON.
The OMICRON power supply DC side negative is connected to the ground
(PE).
5. Remove the
Safety Connector Dongle
from the
SAFETY
connector on the main device and plug it in
the
SAFETY OUT
connector of the
SAA2
control unit. In case of
VOTANO 100
remove
SAB1
connector from
VOB2
SAFETY
output connector.
NOTE!
Alternatively, you can connect another
SAA2
control unit or other
Safety Accessory
as
described in this Supplementary Sheet to the
SAFETY OUT
connector of the control unit
instead of the
Safety Connector Dongle
, in which case you must plug the
Safety Connector
Dongle
to the
SAFETY
OUT
connector of the last cascaded
Safety Accessory
.
Mounting ring
Magnet or
M4 thread
¼“ thread (torque 0.8 Nm)
Tripod mounting
1
1
Tripod is not included in OMICRON product range.