User
Manual
Version
1.0
43
Figure
20:
DPSS
specific
boards
and
connections
Please
turn
on
laser
by
laser
now
and
check
the
output
power.
The
new
laser
head
should
emit
laser
light
in
the
pre
‐
aligned
direction.
It
needs
final
adjustment
to
be
launched
in
the
fibre
coupler,
so
please
refer
to
the
alignment
procedure
described
in
chapter
6.
ATTENTION:
Do
not
look
into
the
beam
or
into
reflections
of
the
beam
on
surfaces
while
adjusting
the
beam
to
the
beam
path!
When
finished
with
the
adjustment
procedure,
please
put
back
the
top
cover
seal
in
its
original
position.
Please
use
the
top
cover
screw
spacers
for
finding
the
right
position
of
the
seal.
After
closing
the
top
cover,
please
re
‐
check
the
alignment
of
the
new
added
laser
beam.
Please
be
careful
with
the
top
cover
seal.
Do
not
use
sharp
tools
to
position
the
top
cover
seal.
The
top
cover
should
not
be
left
open
for
a
long
time.
After
finishing
the
placement
of
the
additional
laser
module,
it
should
be
closed
immediately.
Please
tighten
the
top
cover
screws
until
the
torque
will
be
increased.
There
are
spacers
installed
to
avoid
damage
to
the
seal.
The
screws
cannot
be
screwed
in
more
than
the
spacer
allows.