3.1 Status
LEDs
State
Error
LED
Interlock
LED
Emission
LED
Fibre
Out
1
Fibre
Out
1
Shutter
Status
Colour
Red
Yellow
White
White
White
White
D
EVICE
OFF
OFF
OFF
OFF
OFF
OFF
OFF
D
EVICE
ON
ON
1
/OFF
OFF
OFF²
OFF
OFF
OFF
D
EVICE
L
ASER
ON
ON
1
OFF
ON
ON/OFF
ON/OFF
ON/OFF
D
EVICE
E
RROR
ON
1
OFF
OFF
OFF
OFF
OFF
D
EVICE
I
NTERLOCK
ON
ON
OFF
OFF
OFF
OFF
Table
8:
LED
signals
1
Error
LED
is
blinking,
if
at
least
one
of
the
lasers
inside
the
box
has
a
problem.
Error
LED
is
on,
if
a
system
error
occurs.
²
Emission
LED
is
blinking
if
a
key
switch
toggle
is
needed.
This
is
only
possible
if
device
is
in
CDRH
mode.
See
chapter
3.4.
In
addition
to
the
status
LEDs
in
the
table
above,
the
power
switch
is
backlit
with
a
blue
LED.
This
LED
indicates
several
states
for
the
device.
Please
find
details
in
the
following
table:
Power
LED
(blue)
state/duty
‐
cycle
Power
State
Keyswitch
State
Temperature
Regulation
State
LED
OFF
Device
not
powered,
push
button
to
power
up
or
plug
in
mains
‐‐
‐‐
LED
blinking
with
10%
duty
‐
cycle
Device
powered
Keyswitch
OFF
Device
temperature
not
stable
LED
blinking
with
50%
duty
‐
cycle
Device
powered
Keyswitch
OFF
Device
temperature
stable
LED
blinking
with
90%
duty
‐
cycle
Device
powered
Keyswitch
ON
Device
temperature
not
stable
LED
ON
Device
powered
Keyswitch
ON
Device
temperature
stable
Table
9:
Power
LED
States
If
the
temperature
regulation
state
indicates
a
stable
temperature,
the
device
is
within
±4°C
of
the
setpoint.
Wavelength
and
power
output
is
only
stable,
if
temperature
is
within
this
range.