CMC 310 User Manual
30
OMICRON
4.5
Inputs
4.5.1
Binary inputs
General data of binary inputs 1…6
Number of binary inputs
6
Trigger criteria
Potential-free or DC-voltage compared to threshold voltage.
Reaction time
Max. 220 µs
Sampling rate
10 kHz
Time resolution
100 µs
Maximum measuring time
Unlimited
Debounce/deglitch time
0 … 25 ms (→ page 31)
Counting function
Counter frequency
Pulse width
<3 kHz (per input)
>150 µs (for high and low signals)
Connection
4 mm sockets
Insulation
3 galvanically insulated binary groups with each 2 inputs
having its own GND. Functional insulation to the power
outputs, DC inputs and between galvanically separated
groups. Reinforced insulation from all SELV interfaces and
from power supply.
Data for potential-sensing operation
Range/resolution
20 ... 300 V
0 ... 20 V
500 mV
50 mV
Maximum input voltage
CAT IV: 150 V
CAT III: 300 V
Threshold voltage accuracy
1
1. Valid for positive voltage signal edge; percentage is shown in respect to each range’s full-scale.
5 % of rd. + 0.5 % of rg.
Typical threshold voltage hysteresis
Range 20 ... 300 V: 900 mV
Range 0 ... 20 V: 60 mV
Input impedance
Threshold 20 ... 300 V: 135 kΩ
Threshold 0 ... 20 V: 210 kΩ
Summary of Contents for CMC 310