Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 16 of 21
Smart Measurement Solutions
Smart Measurement Solutions
®
1.8 Check of attenuator correction
This chapter shows how to do a rough check of the attenuator correction.
Measurement setup:
Connect the Source of Bode 100 to the two inputs using the cables and the T-BNC Adapter delivered
with Bode 100.
Figure 36: test attenuators - measurement setup
Equipment setup:
Please select the
measurement type “Gain / Phase” and apply the following settings in the hardware
setup:
Figure 37: test attenuators - measurement type
Figure 38: test attenuators
– measurement settings and hardware setup