General Testing: 

 
1. 

Release the clamps on front door to have access to the test section. 

2. 

Mount your specimen (e.g., component, PCB or heat sink) in the test section. 

3. 

Clamp the door back on the WT-3100. 

4. 

Place flow measurement instrument in the instrument port. 

5. 

Connect the fan cable to the controller box (5). 

6. 

Connect the fan controller box (5) to a DC power supply or wind tunnel controller.             

Important Note:

 Voltage to the fans cannot exceed 26 Volts-DC. 

7. 

Adjust voltage until desired flow rate is attained. 

 

 
 
 
Component and Heat Sink Testing: 

 
1. 

Release the clamps on front door to have access to the test section. 

2. 

Mount your component(s) in the test section. 

3. 

Clamp the door back on the WT-3100. 

4. 

Connect the fans to a variable DC power supply. Voltage to the fans cannot exceed 26 
Volts-DC. Adjust voltage until desired flow level is attained. 

5. 

Insert the velocity and temperature probes into the instrument ports provisioned in 
upstream of the test section. 

6. 

Turn on the power supplies and start the experiment. 

 
Note: If the wind tunnel is provided with a heat slug, please apply a coating of thermal grease to 
the cylindrical heater and place it inside the hole in the Delrin housing behind the test plate. 
Measure the resistance of the heater with a Multimeter. Connect the the heater to an external 
power supply and adjust the voltage for the desired power dissipation. Please make sure that 
the temperature of the heater does not become excessive, otherwise it will damage the Delrin 
hosing. Depending on the heat sink that is mounted on the slug, always start with a low Watt to 
get a feeling for the temperature. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

SYSTEM

 

OPERATION 

Summary of Contents for WT-3100

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info WT 3100 Controlled Open Loop Wind Tunnel Shop online at User sGuide...

Page 2: ...ged by the United Kingdom Office Toll Free 0800 466 342 TEL 33 0 161 37 29 00 FAX 33 0 130 57 54 27 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Toll Free 0 800...

Page 3: ...he front door for mounting of the boards A unique internal rail guides provide mechanisms for installation of test specimen of different sizes e g PCB heat sink Installation of different fan trays can...

Page 4: ...on Variable Speed Change the flow rate by controlling the fan RPM Flow Direction Test the effect of flow direction fan failure simulation by controlling the fans either variable RPM or on off Quick Ac...

Page 5: ...Figure 1 The WT 3100 System Components...

Page 6: ...variable DC power supply Voltage to the fans cannot exceed 26 Volts DC Adjust voltage until desired flow level is attained 5 Insert the velocity and temperature probes into the instrument ports provis...

Page 7: ...uadrant of the test section for two different velocity levels 0 75 m s and 4 m s 150 and 800 ft min respectively VELOCITY PROFILE WT 3100 test section top view Velocity measurement quadrant Velocity D...

Page 8: ...0 40 151 773 40 10 153 796 40 20 153 773 40 30 152 775 40 40 158 771 60 10 154 785 60 20 152 775 60 30 153 775 60 40 152 776 90 10 154 784 90 20 153 768 90 30 154 766 90 40 155 766 120 10 155 780 120...

Page 9: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 10: ...city Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industria...

Reviews: