TROUBLESHOOTING

Once you apply the steps described above, if you still cannot see

the Transmitter on your VC check the following items:

1. Blue LED

– The Blue LED on the Transmitter blinks every time 

it transmits data. If the Blue LED is solid on it then means that 

it’s trying to connect and transmit data to the access point with 

no success.

2. Java Runtime Environment (JRE)

– Make sure that the 

Java Runtime Environment (JRE) is running on the PC on 

which the VC is installed. If the Java Runtime is not installed 

on the computer the VC will not display readings from the 

Transmitters. Should be 32-bit Java version 1.6 or higher.

3. Wireless Connection –

If you plan to connect on a wireless 

LAN make sure that the wireless connection on your computer on

which the VC is installed is linked to the correct access point. You 

may want to connect to the same access point to which the 

Transmitter is connected. You can verify that by looking at the 

Wireless Connection Manager on your computer.

4. Wired LAN

– If your computer is on a wired LAN on which 

the correct access point is also connected, make sure that 

you can reach the access point from your computer. This 

can be accomplished by “pinging” the IP address of the 

access point.

5. Firewall

– Make sure that the firewall is off or the exceptions are 

added properly. 

6. Back to AD-HOC Mode

– If you cannot find anything wrong 

with your computer wireless connection and the access point, 

then there is a chance that the Transmitter was not configured 

correctly. To reconfigure the Transmitter you must put the 

Transmitter back into the AD-HOC mode. To do this, open the 

case, slide the red Power button to the OFF position. Next, press 

and hold the white Default button, slide the red Power switch 

back to the ON position, and release the white button once the 

blue LED comes up solid. You can now follow from 

Step 3.4

of 

the Transmitter’s Initial Configuration section.

7. Access Point/Wireless Router

– In general, the latest/newest 

access Points are better than older ones. Also, check to see if 

your access point has the latest firmware installed.

SPECIFICATION

SENSOR SPECIFICATIONS

RELATIVE HUMIDITY (wTHP, wTHP2, wBTHP)

Accuracy/Range

±2% for 10 to 90%

±3% for 5 to 10% and 90 to 95%

±4% for 0 to 5% and 95 to 100%

Hysteresis

: ±1% RH

Non-linearity

: ±3%

Repeatability

: ±0.1%

Resolution: 

0.1%

TEMPERATURE 

Accuracy/Range*:

wTHP, wTHP2

±0.5°C for 5 to 45°C (±0.9°F for 41 to 113°F); 

up to ±1.5°C for -40 to 5°C and 45 to 124°C 

(up to ±2.7°F for -40 to 41°F and 113 to 255°F)

wTP1, wTP2

±0.5°C for 10 to 85°C (±0.9°F for 50 to 185°F);

±1°C for -40 to 10°C and 85 to 125°C 

(±1.8°F for -40 to 50°F and 185 to 257°F)

wBTHP

±0.5°C for 5 to 45°C (±0.9°F for 41 to 113°F); 

up to ±1.5°C for -40 to 5°C and 45 to 85°C 

(up to ±2.7°F for -40 to 41°F and 113 to 185°F)

wBTP

±0.8°C @ 25ºC (±1.5°F @ 77ºF)

±4°C for -40 to 85°C (±7.2°F for -40 to 185ºF)

*NOTE: extended temp range is for Probe only,

the Controller’s operating temp is -10 to 55ºC
Resolution

: 0.1°C

BAROMETRIC PRESSURE (wBTP, wBTHP)

Accuracy/Range

±2 mbar for 300 to 1100 mbar @ 0 to 50ºC

±6 mbar for 300 to 1100 mbar @ -40 to 85ºC

Resolution

: 0.1 mbar

ANALOG VOLTAGE & CURRENT INPUT (wVI)

Voltage Input: 

Differential; bipolar; ±100 mV, ±1 V, ±10 V

Input Impedance: 

38 K ohm for voltage

Current Input:

Differential; bipolar; ±20 mA (5 ohm load)

Accuracy: 

±0.1% Full Range @ 25ºC

Reading Rate

: Periodic (1 sample/update) or

continuous (3 samples/second)

Resolution

: 16 bits

THERMOCOUPLE INPUT (wTC)

Thermocouple Type (ITS 90)

:

J, K, T, E, R, S, B, C, N, L

Reading Rate

: Periodic (1 sample/update) or

continuous (3 samples/second)

METER SPECIFICATIONS

Supported Protocols (Transmitter)

: TCP/IP,

UDP, ARP, ICMP, DHCP, HTTP and FTP

Supported Protocols (VC)

TCP/IP, UDP, HTTP, FTP, SMTP and Telnet

WIRELESS COMMUNICATION

Standard

: IEEE 802.11 b/g 

Frequency: 

2.4 GHz (2402~2483.5 MHz)

Range: 

60m (200') indoor line-of-site 

or more depending upon sensitivity, 

data rate, wireless access point and

environmental considerations

Radio Power Output Level (Class 1):

91.4 mW EIRP (19.6 dBm EIRP)

Modulation:

802.11b compatibility:

DSSS (CCK-11,

CCK-5.5, DQPSK-2, DBPSK-1);

802.11g:

OFDM (default)

Channels:

1 to 13; Channel 14 for Japan use

only and is not certified.

Channel Spacing (Bandwidth):

20 MHz

Transmission Rate (Over the Air):

802.11b:

1 to 11 Mbps for 

802.11g:

6 to 54 Mbps

POWER (wSERIES AC POWER)

Power Input:

5 Vdc

Consumption:

0.7 W max 

Safety Qualified AC Power Adapter

(included):

Nominal Output:

5 Vdc @ 0.6 A 

Input: 

100 to 240 Vac, 50/60 Hz

Back-up Alkaline Battery:

One AA 1.5 Vdc, supplied

POWER (wSERIES-CCELL)

Alkaline Battery:

Two C-CELL 1.5 Vdc, supplied

Lifetime:

Estimate of 2.4 years (wTC/wVI) and

4.3 years (wTxxP/wBxxP) with frequency of 1

reading per 1 minute

Dimensions:

96.5H x 146.3W x 50.8D mm (3.8 x 5.76 x 2"),

not including connectors or antenna

Operating Temperature:

-10 to 55ºC (14 to 131ºF), 90% RH non-condensing

AC Power Adapter: 

0 to 40ºC (32 to 104ºF)

b)

Channel

– This is the Wi-Fi channel on which the 

access point is accepts connections. If not sure what 

channel to use, simply select Auto.

c)

Authentication

– This is the security method which the 

access point is using for connections. Choose the 

desired security based on the present access point settings.

Use WPA2-PSK for the most secure

connection. No security can be used to

conserve battery power. WPA2-PSK 

Enterprise is not supported

d)

Passphrase

– If the authentication is WPA2-PSK, then 

enter a passphrase here.

• Passphrase restricted characters include 

“ ;

• The minimum characters that can be used for 

Passphrase/WPA2 is 8, maximum is 63.

e)

Default Key

– This is the password if authentication is 

WEP-128.

• WEP KEY can only use the following characters: 

a b c d e f 0 1 2 3 4 5 6 7 8 9

.

• WEP KEY must be a length of 26 characters.

f)

DHCP

– Check this option if the Transmitter will receive 

a Dynamic IP address from a DHCP server.

For initial configuration, it is recommended to

use DHCP for assigning IP Address to the

Transmitter. The IP address can be made

STATIC from the Virtual Coordinator after the

Transmitter starts communicating with it.

g)

IP Address

– Assign a STATIC IP address.

h)

Netmask

– Assign a subnet mask associated with the 

STATIC IP address.

i)

Gateway Address

– This could be the IP address of 

your access point or wireless router which the 

Transmitter is associating with.

j)

Virtual Coordinator IP

– Assign the IP address of the 

PC on which the VC is installed.

k)

Communication Protocol

– This is the protocol used 

to communicate with the VC. Either TCP or UDP 

protocol can be selected. UDP is default and uses less 

battery power.

l)

Virtual Coordinator Port

– Readings from the sensor 

(Transmitter) are sent to the VC using this port number.

Use the default UDP protocol and port 

50002

unless 

this port is already used on the VC. If TCP protocol is 

used change this port number to 50006.

m)

Update Period

– Enter the Transmitter’s reading 

transmission frequency in seconds.

n)

Save and Reboot

– Click “Save Changes”. Verify the 

settings and then click “Reboot”. The webpage should

say that the Transmitter is rebooted.
If your Transmitter has a display, you should now be 

able to see the sensor readings on the display.

The Transmitter stays in AD-HOC mode for a

few minutes and then goes to sleep to 

conserve battery. So the step to put the device

into AD-HOC mode and to configure it using a

web page needs to be done together.

Also during this time, it is recommended to set the 

“Update Period” to the default (10 seconds). Once the 

Transmitter and the VC are communicating properly, the 

“Update Period” can be changed to the desired value. 

3.6. Getting the Readings

If the PC used to configure the Transmitter is the same as the 

one running the VC, then change its IP address to the STATIC 

IP address set for the VC. Start a new browser and enter 

http://<IP address of PC running virtual coordinator>

If using a different PC, go to the PC running VC. Open up a web 

browser and enter 

http://<IP address of virtual coordinator>

Click the button that says readings. Enter the user name as 

user

” and the password as “

12345678

”. You should be able to 

see the readings.

READINGS

http://

192.168.0.105

/reading_html

Figure 8

Series

SETUP GUIDE

Wireless

Transmitter

Wi-Fi

802.11b/g Wireless Ethernet

MQS5029/0717

It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly

pursuing  certification  of  its  products  to  the  European  New Approach  Directives.  OMEGA  will  add  the  CE  mark  to  every

appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for

any errors it contains, and reserves the right to alter specifications without notice.

TRADEMARK NOTICE:

®

,

omega.com

®

,

are Trademarks of OMEGA ENGINEERING, INC.

®

This  device  is  marked  with  the  international  caution  symbol.  It  is  important  to  read  the

Setup Guide before installing or commissioning this device, as the guide contains important

information relating to safety and EMC.

WARNING:

These products are not designed for use in, and should not be used for, patient-connected applications.

omega.com   [email protected]

®

 

Servicing North America:

U.S.A.  

Omega Engineering, Inc. 

Headquarters:  

Toll-Free: 1-800-826-6342 (USA & Canada only) 

 

Customer Service: 1-800-622-2378  

 

(USA & Canada only) 

 

Engineering Service: 1-800-872-9436  

 

(USA & Canada only) 

 

Tel: (203) 359-1660 

Fax: (203) 359-7700 

 

e-mail: [email protected] 

 

 

 

For Other Locations Visit omega.com/worldwide

RoHS 2 Compliant

OMEGA’s  policy  is  to  make  running  changes,  not  model  changes,  whenever  an  improvement  is  possible.  This  affords  our 

customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
©  Copyright  2017  OMEGA  ENGINEERING,  INC.  All  rights  reserved.  This  document  may  not  be  copied,  photocopied, 

reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior 

written consent of OMEGA ENGINEERING, INC.

FOR 

WARRANTY

 RETURNS, please have the  

following information available BEFORE contacting 

OMEGA:
1.   Purchase Order number under which the product 

was PURCHASED,

2.   Model and serial number of the product under 

warranty, and

3.  Repair instructions and/or specific problems  

  relative to the product.

FOR 

NON-WARRANTY

 REPAIRS, consult 

OMEGA for current repair charges. Have 

the following information available BEFORE 

contacting OMEGA:
1.  Purchase Order number to cover the COST  

  of the repair,
2.  Model and serial number of the product, and
3.  Repair instructions and/or specific problems  

  relative to the product.

RETURN REQUESTS/INQUIRIES

Direct  all  warranty  and  repair  requests/inquiries  to  the  OMEGA  Customer  Service  Department.  BEFORE 

RETURNING  ANY  PRODUCT(S)  TO  OMEGA,  PURCHASER  MUST  OBTAIN  AN  AUTHORIZED  RETURN  (AR) 

NUMBER  FROM  OMEGA’S  CUSTOMER  SERVICE  DEPARTMENT  (IN  ORDER  TO  AVOID  PROCESSING 

DELAYS). The assigned AR number should then be marked on the outside of the return package and on any 

correspondence.
The  purchaser  is  responsible  for  shipping  charges,  freight,  insurance  and  proper  packaging  to  prevent 

breakage in transit.

WARRANTY/DISCLAIMER

OMEGA  ENGINEERING,  INC.  warrants  this  unit  to  be  free  of  defects  in  materials  and  workmanship  for  a 

period of 

13 months

 from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace 

period to the normal 

one (1) year product warranty

 to cover handling and shipping time. This ensures 

that OMEGA’s customers receive maximum coverage on each product.
If  the  unit  malfunctions,  it  must  be  returned  to  the  factory  for  evaluation.  OMEGA’s  Customer  Service 

Department  will  issue  an  Authorized  Return  (AR)  number  immediately  upon  phone  or  written  request. 

Upon  examination  by  OMEGA,  if  the  unit  is  found  to  be  defective,  it  will  be  repaired  or  replaced  at  no 

charge.  OMEGA’s  WARRANTY  does  not  apply  to  defects  resulting  from  any  action  of  the  purchaser, 

including  but  not  limited  to  mishandling,  improper  interfacing,  operation  outside  of  design  limits,  

improper  repair,  or  unauthorized  modification.  This  WARRANTY  is  VOID  if  the  unit  shows  evidence  of  

having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; 

or  current,  heat,  moisture  or  vibration;  improper  specification;  misapplication;  misuse  or  other  operating 

conditions  outside  of  OMEGA’s  control.  Components  in  which  wear  is  not  warranted,  include  but  are  not  

limited to contact points, fuses, and triacs.

OMEGA  is  pleased  to  offer  suggestions  on  the  use  of  its  various  products.  However,  

OMEGA  neither  assumes  responsibility  for  any  omissions  or  errors  nor  assumes  liability  for 

any damages that result from the use of its products in accordance with information provided 

by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the 

company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR  

REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF 

TITLE,  AND  ALL  IMPLIED  WARRANTIES  INCLUDING  ANY  WARRANTY  OF  MERCHANTABILITY 

AND  FITNESS  FOR  A  PARTICULAR  PURPOSE  ARE  HEREBY  DISCLAIMED.  LIMITATION  OF  

LIABILITY:  The  remedies  of  purchaser  set  forth  herein  are  exclusive,  and  the  total  liability  of  

OMEGA  with  respect  to  this  order,  whether  based  on  contract,  warranty,  negligence,  

indemnification,  strict  liability  or  otherwise,  shall  not  exceed  the  purchase  price  of  the  

component  upon  which  liability  is  based.  In  no  event  shall  OMEGA  be  liable  for  

consequential, incidental or special damages.

CONDITIONS:  Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic 

Component”  under  10  CFR  21  (NRC),  used  in  or  with  any  nuclear  installation  or  activity;  or  (2)  in  medical 

applications  or  used  on  humans.  Should  any  Product(s)  be  used  in  or  with  any  nuclear  installation  or 

activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility 

as  set  forth  in  our  basic  WARRANTY/DISCLAIMER  language,  and,  additionally,  purchaser  will  indemnify 

OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the 

Product(s) in such a manner.

Reviews: