background image

If the primary readout shows 

0.0%

with the meter in your hand (and not touching anything), the

meter is ready to make measurements. To measure the moisture level of a material, press the

pinless sensor on the back of the meter (Fig. 1, Callout L) against it. The moisture level, as

percentage, will appear on the primary readout.
If the primary readout shows a value other than 0.0%, you should reset the baseline of

measurements in 

MOIST INT

mode. To do so, continue to hold the meter in your hand (touching

nothing) and press the 

ZERO

button. Immediately after the term 

ZERO

appears briefly below the

bar graph at the left of the display, the primary readout will show 

0.0%

and the measurements

will again be accurate in 

MOIST INT

mode. 

To obtain accurate readings in 

MOIST INT

mode, you should reset

the readout to 0.0% each time you re-enter that mode from 

another mode.

CHECKING CALIBRATION

Although you cannot calibrate the RH650 yourself,  you can 

check 

the meter’s calibration 

periodically to assure yourself of the accuracy of humidity-related readings (RH, absolute

humidity, dew point and mixing ratio). Factory calibration should minimize the humidity sensor’s

drift for several years—well beyond the warranty period.
At a minimum, a calibration check should entail immersing the slotted structure protecting the

humidity sensor in two bottles containing saturated salts. One reference salt should produce a

reading of a 33%; the other should produce a reading of 75%.
Depending on the size and shape of the calibration salt bottle that you use, you may have to

remove the hanger tab (Fig. 1, Callout B) from the meter to provide sufficient clearance for the

bottle to fit snugly over the slotted structure protecting the humidity/temperature sensor (after

removing the dust cap). The hanger tab is removable for that reason.

12

Summary of Contents for RH650

Page 1: ...RH650 9 In 1 Thermo Hygrometer Moisture Meter e mail info omega com For latest product manuals omegamanual info Shop online at omega comSM User sGuide...

Page 2: ...notice Servicing North America U S A Omega Engineering Inc One Omega Drive P O Box 4047 Stamford CT 06907 0047 USA Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada o...

Page 3: ...9 IN 1 THERMO HYGROMETER MOISTURE METER USER S MANUAL RH650 Please read this manual carefully and thoroughly before using this product...

Page 4: ...restoration contractors HVAC R system installers and technicians facility maintenance professionals and electronics manufacturers The first three groups will exploit the meter s ability to measure di...

Page 5: ...terial s moisture level and its dielectric constant and therefore its capacitance 2 The so called fringing field effect the slight spreading of the electric field produced by current flowing between t...

Page 6: ...ecalls 20 pairs of readings Settable high and low alarms for RH relative moisture level and absolute moisture level Backlit LCD with 3 digit primary readout 4 digit secondary readout Zero reset in pin...

Page 7: ...ameters Ambient Temperature Dew Point F or C Absolute Humidity mg L or g m3 Mixing Ratio GPP or g kg Barometric Pressure and Vapor Pressure mBAR or KPa In memory mode used to clear stored readings Als...

Page 8: ...mode indicator Pinless moisture measurement mode indicator Alarm limit indicators Zero reset indicator Pin pinless moisture level indicator Maximum minimum value indicators Data hold indicator Primary...

Page 9: ...conds The shutdown will be announced by two short beeps After powering off the meter replace the sensor protection cap if you do not expect to use the instrument within the next few hours By default t...

Page 10: ...never any button is pressed It will remain on for 30 seconds and then extinguish to extend battery life Holding readings Briefly pressing the ON OFF HOLD button freezes both the primary and secondary...

Page 11: ...e DN button to recall them in reverse order Pressing the UP button when the last stored pair of readings is displayed returns the Record number indicator to 01 corresponding to the first pair of readi...

Page 12: ...OW setpoint as explained earlier in the paragraph on HIGH setpoints Alternatively you can leave the LOW setpoint unchanged Whether or not you make a change your next step must be to press the STORE EN...

Page 13: ...aterial In MOIST EXT mode the meter uses the included pin type probe to measure the conductivity of a material as a proxy for its moisture content In both moisture level measurement modes the RH650 s...

Page 14: ...0 each time you re enter that mode from another mode CHECKING CALIBRATION Although you cannot calibrate the RH650 yourself you can check the meter s calibration periodically to assure yourself of the...

Page 15: ...to 99 9 Pinless Moisture Measurement Depth 0 75 in 19mm Dew Point Range 76 to 140 F 60 to 60 C Atmospheric Pressure Range 30 to 120 kPA Vapor Pressure Range 0 to 20 0 kPA Sampling Rate 2X sec Memory C...

Page 16: ...disassemble the unit Opening the case voids the warranty To avoid fouling the humidity temperature sensor do not operate the unit in very dusty environments Replace the dust cap protecting the sensor...

Page 17: ...per specification misapplication misuse or other operating conditions outside of OMEGA s control Components in which wear is not warranted include but are not limited to contact points fuses and triac...

Page 18: ...bine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industri...

Reviews: