Calibrating the pH electrode

 

The optional pH electrode you are using must be calibrated before making pH measurements 
with it for the first time. The electrode also should be calibrated whenever you wish to maximize 
the accuracy of measurements. Calibration is necessary because not every pH electrode is an 
“ideal” electrode that generates 0 mV at pH 7 and 177.4 mV at pH 4. 
 

1.

 

To calibrate the pH electrode

, perform the following procedure while referring to 

Figure 2. 

 

 

2. Calibrating a pH electrode using the buffer solution  

 
2-1 Electrode plug 
2-2 Electrode handle 
2-3 Electrode head 
2-4 pH buffer solution 
 

1.

 

Insert the plug of the pH electrode

 into the BNC socket at the top of the meter (callout 

1-14 of Fig. 1). 

2.

 

Power on the meter

 by pressing the POWER button. 

3.

 

Using the procedure detailed earlier, 

manually adjust the compensation temperature

 

so it matches the temperature of the buffer solution. Alternatively, use an ATC 
temperature probe to make the adjustment automatically. 

4.

 

Holding it by its handle, 

immerse the pH electrode

 in the bottle of buffer solution. (If 

you are using the ATC temperature probe, immerse its head in the bottle as well.) 

5.

 

Gently shake the electrode

 (and the probe, if used). The display will show the pH value. 

6.

 

Use two fingers to press the REC/ENTER and HOLD/ESC buttons

 at the same time. 

When this screen 

 appears, release the buttons. 

5

Summary of Contents for PHH-SD1

Page 1: ...PHH SD1 pH Meter With Data Logging SD Card e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide NORWALK CT TM...

Page 2: ...pecifications without notice omega com info omega com Servicing North America U S A Headquarters Omega Engineering Inc Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Page 3: ...es 1 Operating Instructions 2 What s in the case 2 Setup 3 Calibrating the pH electrode 5 Normal operation 6 Holding and storing measurements 7 Automatic vs manual datalogging 7 Transferring data from...

Page 4: ...2 pH units 2 digits for pH readings and 0 5 of the reading 2 digits for ORP readings Because it is microprocessor based the PHH SD1 can make full use of the portability reliability and large storage c...

Page 5: ...the front back top bottom and right side of the PHH SD1 Familiarize yourself with the positions and functions of these controls indicators and connectors before moving on to the setup procedure 1 The...

Page 6: ...he screen will revert to the normal display To return to the setup sequence press and hold the LOGGER SET button again To move ahead to the next field in the sequence or to the next parameter press th...

Page 7: ...ss the button until the desired setting yES or no is displayed and then press the REC ENTER button to store the selection 8 Choose a decimal point or comma to represent the decimal division between in...

Page 8: ...d 2 4 pH buffer solution 1 Insert the plug of the pH electrode into the BNC socket at the top of the meter callout 1 14 of Fig 1 2 Power on the meter by pressing the POWER button 3 Using the procedure...

Page 9: ...7 pH 4 and pH 10 Normal operation The PHH SD1 can measure either the pH or the oxidation reduction potential ORP of a solution 1 To select the type of measurement press the button which has the word...

Page 10: ...e TIME button Doing so causes both values to appear briefly at the lower left of the display Holding and storing measurements 4 To hold a measured value of pH or ORP press the HOLD ESC button during t...

Page 11: ...his mode the instrument is storing measurements and their time stamps on the SD card In manual datalogging mode however measurements are being stored continuously with a sampling time of zero and thei...

Page 12: ...3 Typical Excel data only screen 4 Typical Excel graphics only screen 5 Typical Excel mixed data graphics screen 9...

Page 13: ...normal operation with backlight off and SD card not saving data 37 mADC with backlight on and card saving data 49 mADC with backlight on and card saving data Dimensions 6 97 x 2 68 x 1 77 inches 177 x...

Page 14: ...ecification misapplication misuse or other operating conditions outside of OMEGA s control Components in which wear is not warranted include but are not limited to contact points fuses and triacs OMEG...

Page 15: ...s M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U...

Reviews: