5.7 P07
:
ALARM RELAY 3 SET-UP SUB-FUNCTION
1.
Select P07 sub function, Press
to confirm. Enter into concrete set-up procedure.
2.
Select concrete working mode. Press
or
to select suitable control mode.
–
CLE
=
relay 3 as cleaning control relay
–
CAL
=
relay 3 as calibrating indication relay
–
ALL
=
move with two relays simultaneously
–
SP1
=
relay 3 moves with SP1 (limited control mode only)
–
SP2
=
relay 3 moves with SP2 (limited control mode only)
–
OFF
=
relay 3 is off
Press
ENTER
to confirm your set up.
If user select CAL working mode, he can set calibration indicate intervals in the function
form, use houras the unit. In this way, instrument and electrode can keep their reliability.
If user select CLE working mode, he can set cleaning time intervals in the function form,
use hour as the unit; cleaning continuous time, use second as the unit. In this way,
instrument and electrode can keep their reliability. User can install cleaning equipment to
increase service life and reliability of electrode.
3.
Press
to confirm and return to P07 sub-function form. Use
or
to
select other sub-functions
to set up.
NOTE
:
Press
to quit set up mode at any time. Instrument will return to
measurement mode automatically.
done
done
done
HOLD
SET
HOLD
SET
HOLD
SET
Sec
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
Sp2
SET
HOLD
Sp
SET
HOLD
SET
HOLD
ESC
ESC
ENT
ENT
ESC
ENT
ENT
ENT
ENT
ENT
ENT
ENT
7.00
SPI
ATC
O
C
SP2
I2.00 25.0
MEA
PH
7.00
SPI
ATC
O
C
SP2
I2.00 25.0
MEA
PH
7.00
SPI
ATC
O
C
SP2
I2.00 25.0
MEA
PH