Calibrate the Controller:

1.

If it’s not already shown, press 

MENU

until the

controller displays 

r

rd

d..c

cf

f

.

2.

Press 

T/mV

.  The controller

displays 

r

r..1

 

=

2

2

or 

r

r..1

 

=

3

3

(2 or 3 point calibration).

3.

Press 

/pH

to select the

desired calibration type.

4.

Press 

T/mV

to display 

r

r..2

 

=

f

f

or 

r

r..2

 

=

c

c

(temperature unit of F or C).

5.

Press 

/pH

to select the

desired temperature unit.

6.

Press 

MENU

to select the calibration.  The unit

displays.

7.

Proceed with Calibration depending on
Temperature Compensation Setting:

If you choose OFF, go to “To Perform 
Calibration” section.

If you choose METR, ensure the RTD is 
properly connected and go to “To Perform 
Calibration” section.

If you choose MANU, configure temperature 
as described in the next section.

T/mV

MENU

RESET

/pH

SETPTS

T/mV

MENU

RESET

/pH

SETPTS

SETPTS

T/mV

MENU

RESET

/pH

=

Connect the pH  Electrode

1.

Secure the pH electrode to the pH input BNC
connector on the back of the controller.

2.

If you are using automatic temperature
compensation (ATC), connect the RTD as shown
below.

Apply Power

1.

Apply ac power to the unit.

The unit initializes, flashing the following
messages: 

r

rs

st

t

p

ph

h

i

in

ni

it

t

.  Then a pH value

appears.

2.

Verify that a value appears. If not:

Remove ac power.

Verify the TB1 power connections.

Check your power source.

Apply ac power again.

TB1

TB2

1 2

3 4 5 6

7 8

9 10

1 2

3

11 12

RTD

+ -

TB1

TB2

1 2

3 4 5 6

7 8

9 10

1 2

3

11 12

+ -

pH Input

pH Electrode

START HERE

2

3

4

Using This Quick Start Manua

l

Use this Quick Start manual to set up your pH
Controller and begin operation.  Information is provided
on how to:

Mount the controller

Connect ac Power

Connect a pH electrode

Calibrate the controller prior to use

For complete information on this controller, refer to the
Operator’s Manual.

Before You Begin

In addition to the meter and the related parts, you will
need the following items to set up your meter:

ac power, as listed on meter’s ID/Power Label

pH electrode (with BNC input connector)

1

8

” flat blade screwdriver

Mount the Unit

1. Cut a panel opening using

the dimensions shown to
the right.

2. Position the unit in the

opening, making sure the
front bezel gasket is flush
to the panel.

3. Slide on mounting bracket to secure.

Connect 

~

ac Power

Warning: Do not connect AC power to your device until you
have completed all input and output connections. This
device must only be installed by a specially trained
electrician with corresponding qualifications. Failure to
follow all instructions and warnings may result in injury!

1. Remove the panel at the back of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each terminal as shown in the

following figure and tighten the lockdown screws.

4. Tug gently on the wires to verify the connections.

To Select Temperature
Compensation Mode:

1.

Press 

MENU

. The controller

displays:

2.

Press 

T/mV

to display the

current setting.

3.

Press 

/pH

to change to the

desired setting. Choices are:

OFF = the controller uses a
constant 25ºC for temperature
compensation.

MANU = the controller uses a manually entered
value for temperature compensation.

METR = the controller uses the RTD input for
automatic temperature compensation.

4.

Press 

MENU

to select the temperature

compensation setting shown.  The controller
displays.

To Set The Decimal Point Position:

1.

If it’s not already shown, press

MENU

until the controller

displays 

d

de

ec

c..p

p

.

2.

Press 

T/mV

.  The controller

displays:

3.

Press 

/pH

to move the decimal point to the

desired location.  The choices are

f

ff

f..f

ff

f

f

ff

ff

f..f

f

, or

f

ff

ff

ff

f

.

4.

Press 

MENU

to select the decimal point position

shown.  The controller displays:

T/mV

MENU

RESET

/pH

SETPTS

T/mV

MENU

RESET

/pH

SETPTS

SETPTS

T/mV

MENU

RESET

/pH

T/mV

MENU

RESET

/pH

SETPTS

T/mV

MENU

RESET

/pH

SETPTS

T/mV

MENU

RESET

/pH

SETPTS

T/mV

MENU

RESET

/pH

SETPTS

T/mV

MENU

RESET

/pH

SETPTS

=

45,00  + 0,61/-0,00

(1.772  + .024/–.000)

92,00  + 0,81/–0,00

(3.622  + .032/–.000)

PANEL THICKNESS

1,5

R(.06)

4 PLCS

6,4 (.25) MAX
0,8 (.03) MIN

NOTE: Dimensions in Millimeters (Inches)

TB1

TB2

1 2

3 4 5 6

7 8

9 10

1 2 3

11 12

FUSE

 EARTH GROUND

GREEN WIRE

 AC Power

Check for proper Earth grounding in the power 
distribution system (single phase).

SWITCH

LINE

NEUTRAL

EARTH

(L) LINE

(N) NEUTRAL

pH INPUT