26
To avoid damaging the pressure module from overpressure, never
apply pressure above the rated maximum printed on the module.
To avoid damaging the pressure module from corrosion, use it only
with specified materials. Refer to the pressure module documentation
for material compatibility.
To measure pressure, follow these steps:
1. Connect the pressure module to the calibrator as shown in Figure
18.
The calibrator can measure pressure on both the upper and the
lower display. This makes it possible to measure pressure in two
different units at the same time.
Note
: Make sure the calibrator is on before you plug in the pressure
module.
2. Switch to either upper or lower display from the Main Menu.
3. Select [PRESSURE] from the primary parameters.
4. Select the desired measuring unit.
5. Zero the pressure module. The zero function on the calibrator can
be found in the pressure zeroing menu.
Figure 18. Measuring Pressure Transmitter
Note:
On high pressure modules engineering units normally
associated with low pressure ranges such as, inH
2
O, cmH
2
O, etc. are
not valid selections. Selecting one of this units with a high pressure
module attached will cause the display to read "----".
Summary of Contents for PCL1200