background image

SPECIFICATIONS

1 - General

Temperature Range

0 to 1000°F (-18 to 538°C)

Accuracy @ 22°C (72°F)

3% of Rdg or 4°F whichever is greater

ambient & Emissivity >= 0.95

Repeatability

1% of Rdg.

Optical Field of View

6 to 1 (Distance to Spot Size)

Spectral Response

5 to 14 microns

Response Time

150 msec, 0 to 63% of final value

Emissivity Range

0.1 to 1.00 , Adjustable

Analog Output

K type Thermocouple, compensated

Output Connection

Universal Type K female socket accepts both
Miniature (SMP) and standard (OST) type 
connectors

Power

3.5 VDC, AA Size Lithium Battery - 1900 mAH

Battery Life

11 Days, continuous (No Laser)
6 Days, continuous with Laser

Power Indication

Flashing red LED, twice per second

Low Battery Indication

Flashing red LED, twice per second

Operating Ambient Temperature

0 to 50°C (32 to 122°F)

Operating Relative Humidity

Less than 95% RH, non-condensing

Dimensions

117 L x 47 H x 27 D mm (4.6" L x 1.84" H x 1.07" D)

Weight

82 g (0.180 lbs)

2- Laser Sighting

Laser Wavelength (Color)

650 nm (Red)

Operating Distance

Up to 7.6 m (25 feet)

Laser Sighting Type

Built-in, Laser circle with a Laser dot in the center

Max. Laser Power Output

Less than 1 mW

Safety Classification

Class 2

FDA Classification

Class II Laser Product

Laser Beam Diameter

Less than 5 mm

Beam Divergence

Less than 2 mrad

Laser Power Switch

Laser key - Momentary switch

Laser power Indication

Flashing red LED, twice per second

Laser Warning

On the front of the unit

& Certification Label

Identification & Laser

On the back of the unit

Aperture Label

OSM101

5

Summary of Contents for OSM101

Page 1: ...rtable Infrared To Analog Converter Module MADE IN Built In Patented Laser Sighting Circle Dot Combined TM omega com e mail info omega com For latest product manuals omegamanual info Shop online at Us...

Page 2: ...n Germany TEL 49 0 7056 9398 0 FAX 49 0 7056 9398 29 Toll Free in Germany 0800 639 7678 e mail info omega de United Kingdom One Omega Drive River Bend Technology Centre ISO 9002 Certified Northbank Ir...

Page 3: ...quipment digital meters chart recorders etc which have no accessible live parts Such equipment should comply with all the applicable safety requirements Do not operate the equipment in flammable or ex...

Page 4: ...flash Connect the thermocouple output of the OSM101 to a thermocouple meter using the retractable cable provided There is a momentary Laser button on the unit Press and hold the laser key The laser LE...

Page 5: ...FICATION LABEL POWER SWITCH EMISSIVITY ADJUST LASER APERTURE LINE OF SIGHT OF THE LASER BEAM Dimensions in mm inches LINE OF SIGHT OF FIELD OF VIEW BUMPER GUARD 15 24 60 19 0 75 101 6 4 0 41 40 1 63 2...

Page 6: ...101 E E F F D D B C B B B B B B B C C H GGG H G H GG H G H G G H G G G H H GG G H H H H G G G LASER APERTURE IDENTIFICATION LABEL BATTERY COMPARTMENT 43 74 1 84 MOUNTING SCREWS OUTPUT Dimensions in mm...

Page 7: ...LED twice per second Low Battery Indication Flashing red LED twice per second Operating Ambient Temperature 0 to 50 C 32 to 122 F Operating Relative Humidity Less than 95 RH non condensing Dimensions...

Page 8: ...wrought iron 0 94 Iron and steel excluding stainless rusted iron plate 0 69 Iron and steel excluding stainless polished steel 0 07 Iron and steel excluding stainless polished steel oxidized at 600 C...

Page 9: ...oda 0 85 0 95 Marble polished light gray 0 93 Paints lacquers and varnishes Black matte shellac 0 91 Paints lacquers and varnishes aluminum paints 0 27 0 67 Paints lacquers and varnishes flat black la...

Page 10: ...NOTES OSM101 8...

Page 11: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 12: ...ty Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: