5.

Record the temperature from the display.

6.

Release the MEASURE button.  The last reading will
remain locked on the display for 10 seconds after
releasing the button.

2.6 SCAN Mode Temperature Measurements

Use this measurement mode for thermal scanning of a tar-
get or when measuring targets located at a distance.  The
OS950 is a wide-angle thermal scanner with a fixed emis-
sivity of 0.9.  The target's temperature is continuously dis-
played and updated 10 times per second for as long as you
hold the MEASURE button down. 

The OS950 displays the average temperature of everything
within the sensor's field of view measurement area.  The
size of the measuring area depends on the distance of the
device from the target.  The OS950 has a 1:1 field of view,
which

means that this unit will measure a 1"diameter target area
when the device is 1" from the target.

2.6.1 General Operation

For continuous temperature measurement of a target or for
temperature measurement at a distance, use the following
procedure.

1.

Determine the appropriate measurement distance from
your desired target.

2.

Make sure the MODE switch is on SCAN.

3.

Point the nosepiece at the target.

4.

Press and hold the MEASURE button.

5.

Record the temperature from the display.

6.

Release the MEASURE button.  The last reading will
remain locked on the display for 10 seconds after
releasing the button.

Chapter Three - Measurement Tips

3.1 Optimizing Target Area Measurements

When performing a continuous thermal scan of a target, or
measuring a target located at a distance, proper target
alignment will help assure accurate results.  Since the
OS950's sensor measurement area is fixed at a field of
view of 1 to 1, it is important to properly determine the best
measuring distance from the target, to optimize the reading.
A field of view of 1 to 1 means that the OS950 viewing area
is 1" diameter for 1" distance from target.  Optimize optical
alignment, so that the OS950 views only the target and
does not overlap the measurement area.  See Figure below
for examples of proper optical alignment.  

5

POOR

POOR

GOOD

Summary of Contents for OS950 Series

Page 1: ...www omega com e mail info omega com omega com User sGuide OS950 SERIES Handheld Infrared Thermometers Shop online at ...

Page 2: ...30 621 400 FAX 33 0 130 699 120 Toll Free in France 0800 4 06342 e mail sales omega fr Germany Daimlerstrasse 26 D 75392 Deckenpfronn Germany Austria Tel 49 0 7056 9398 0 FAX 49 0 7056 9398 29 Toll Free in Germany 0800 639 7678 e mail info omega dl United One Omega Drive River Bend Technology Centre Kingdom Northbank Irlam Manchester ISO 9002 M44 5BD United Kingdom Certified Tel 44 0 161 777 6611 ...

Page 3: ......

Page 4: ......

Page 5: ... 5 MAX Mode Temperature Measurements 4 2 5 1 General Operation 4 2 6 SCAN Mode Temperature Measurements 5 2 6 1 General Operation 5 Chapter Three Measurement Tips 5 3 1 Optimizing Target Area Measurements 5 3 2 Measuring Metallic Targets 6 3 3 Optimizing AESC Measurements for NIST Traceability 6 3 4 Measuring Transparent Materials 6 3 5 Measuring Wet or Icy Surfaces 6 Chapter Four Maintentance 7 4...

Page 6: ...lications such as evaluating insulation monitoring machinery and inspecting electrical equipment Or the OS950 may be operated in the MAX mode where the high est temperature measured while monitoring is presented This manual provides complete instructions on how to oper ate and use your OS950 to provide fast accurate reliable temperature measurements for your application needs 1 2 Product Features ...

Page 7: ... the screw from the underside of the case 2 Remove the cover exposing the battery area 3 Locate the push button switch as shown in the picture 4 With the unit off display is blank push the push but ton in 5 Holding the push button in turn the unit on with the MEASUREMENT button LED Display MEASUREMENT button LED Display Sensor Window AECS Cup Battery Cover Screw Displays target temperature and dia...

Page 8: ...1 General Operation CAUTION SHOCK HAZARD DO NOT TOUCH THE INSTRUMENT TO ELECTRICAL CONDUCTORS The OS950 measures temperature very quickly To obtain an accurate measurement it is only necessary to touch the surface for a fraction of a second For exact surface temperature measurements that are NIST traceable use the following procedure 1 Press and hold the MEASURE button 2 The display will briefly s...

Page 9: ...may be a few degrees too high Allow a few minutes between readings for the nosepiece to cool down and stabilize If the OS950 makes contact with a hot surface and an error message HI shows on the display remove the instrument from the object immediately 2 5 MAX Mode Temperature Measurements Use this measurement mode when making contact with the target or when the target is large The OS950 is a wide...

Page 10: ...t or for temperature measurement at a distance use the following procedure 1 Determine the appropriate measurement distance from your desired target 2 Make sure the MODE switch is on SCAN 3 Point the nosepiece at the target 4 Press and hold the MEASURE button 5 Record the temperature from the display 6 Release the MEASURE button The last reading will remain locked on the display for 10 seconds aft...

Page 11: ...ake sure that the nosepiece is flush with the target when possible so that the reflective cone can complete trap out any light or incident radiation from the environment 3 4 Measuring Transparent Materials Although transparent materials pass visible light many visi bly transparent materials are opaque in the infrared and can be directly measured by the OS950 To test for infrared transparency touch...

Page 12: ...indow and reflective cup with a cotton swab dipped in alcohol Periodic cleaning is a good practice 4 3 Storage When not in use store the OS950 in a container to protect the unit from dirt and dust Do not store the unit in areas of extreme heat cold or humidity If not using the OS950 for a long period remove the battery from the unit 4 4 Calibration The OS950 is factory calibrated under high qualit...

Page 13: ...medy Measurement Fault High Ambient Temperature HI A Ambient 122 F 50 C Move to alternate location Low Ambient Temperature LO A Ambient 32 F 0 C Move to alternate location Low Battery N A Battery power low Replace battery Dead Battery Dead battery Replace battery Other Errors Condition Error Code Remedy Measurement Fault Processing Error E or Err Internal fault Repeat measurement if error remains ...

Page 14: ...he instrument s nosepiece in contact with the surface Touch the nosepiece flush to the target s surface and repeat the measurement If after applying all of the above solutions the instrument s accuracy is still in question contact Omega for further help Chapter Six OS950 Specifications Target Temperature Range OS951 50 to 550 F 45 to 287 C OS953 0 to 1000 C 18 to 538 C OS955 186 to 1207 F 86 to 65...

Page 15: ...9 mm Diameter Spot Area in AECS Mode Spectral Sensitivity 2 to 20 microns Battery One 9 volt Battery Life 5000 readings 10 seconds each Instrument Ambient Temperature Operating Range 32 to 122 F 0 to 50 C Storage Range 22 to 140 F 30 to 60 C Extended storage at temperature extremes will reduce bat tery life 10 ...

Page 16: ......

Page 17: ...e shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans Should ...

Page 18: ...ators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipment DATA ACQUISITION Data Acquisition Engineering Software Communications Based Acquisition Systems Plug in Cards for Apple IBM Compatibles Datalogging Systems Recorders Printers Plotter...

Reviews: