Infrared Thermometer  

 

 

V. OPERATION INSTRUCTIONS 

• 

Measuring Trigger 

To measure a temperature, point the unit at an object and 
pull the trigger. Be sure to consider distance-to-spot size 
ratio and field of view. The laser is used for aiming the 
target for reference. The temperature reading will be 
updated on the LCD. When the operator releases the 
trigger, the reading will automatically be held on the LCD 
for 10 more seconds. After 10 seconds this thermometer 
will power down itself to save battery. 

• 

Distance to Spot Ratio (D/S) 

You have to make sure that the detection area you want 
to measure is larger than the required spot size to get a 
correct reading. The temperature you get is an average 
temperature of the detected region. The smaller the target, 
the shorter distance is required for the measuring. 
(Please refer to the diagram on the side of the unit) 
 

 

               

 

 

• 

Emissivity 

Emissivity is a term used to describe the energy-emitting 
characteristics of materials. The higher of this value 
means the more capability of radiation emittance the 
materials has. Most organic materials and painted or 
oxidized surfaces have an emissivity of 0.98. Metal 
surfaces or shiny materials has a lower emissivity and 

400

Measuring Distance

Unit : mm

800

 Distance : Spot = 8 : 1

52

22

200

113

Spot Size

Summary of Contents for OS542

Page 1: ...www omega com e mail info omega com User sGuide OS542 Infrared Thermometer Shop online at CAUTION This product is not intended for medical use or use on humans ...

Page 2: ...se shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans Should...

Page 3: ...age I SAFETY INFORMATION 2 II CAUTIONS 2 III GENERAL SPECIFICATION 3 IV NAME AND FUNCTION 4 V OPERATION INSTRUCTIONS Measuring Trigger 5 Distance to Spot Ratio D S 5 Emissivity 5 Maintenance 6 Replace Battery 6 ...

Page 4: ...n CE This instrument conforms to the following standards EN50081 1 1992 Electromagnetic Emissions EN50082 1 1997 Electromagnetic Susceptibility Test were conducted using a frequency range of 80 1000 MHz with the instrument in three orientations The average error for the three orientations is 2 0 C 4 0 F at 3V m throughout the spectrum However between 300 MHz and 500 MHz at 3V m the instrument may ...

Page 5: ... SPECIFICATION Display 4 digit LCD Display Illumination Backlight by LED Field of View 8 1 Target Indicator Laser spot Emissivity 0 98 Power Off Automatic power off after 10 seconds Temperature Range 20 500 C 4 932 F Accuracy 2 C 4 F or 2 of reading Resolution 0 5 C 0 5 F Repeatability Within 1 of reading or 1 C 2 F Storage Condition 10 60 C 14 140 F 75 RH Operating Condition 0 40 C 32 104 F 10 90...

Page 6: ...0 670mm SCAN 1 2 6 5 7 F OFF LASER ON C 3 4 CAUTION 1 Laser emitting hole 2 Infrared sensor aperture 3 Laser on off switch 4 C F switch 5 Measuring Trigger 6 Battery cover 7 LCD display LCD Display HOLD SCAN Low battery Temperature unit indication Data Hold indication Laser on indication Digital value indication Scan indication ...

Page 7: ... to make sure that the detection area you want to measure is larger than the required spot size to get a correct reading The temperature you get is an average temperature of the detected region The smaller the target the shorter distance is required for the measuring Please refer to the diagram on the side of the unit Emissivity Emissivity is a term used to describe the energy emitting characteris...

Page 8: ...t use solvents to clean the lens Replace Battery When the battery voltage drops below that required for reliable operation the low battery symbol will appear indicating it is the time to replace the battery To change the 9V battery detach the battery compartment cover by pushing the engraved pattern on the handle and pull down Change the 9V battery with new one and put the battery cover back LASER...

Page 9: ... Guyancourt France Tel 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 0 7056 9398 0 FAX 49 0 7056 9398 29 Toll Free in Germany 0800 639 7678 e mail info omega de United Kingdom One Omega Drive River Bend Technology Centre ISO 9002 Certified Northbank Irlam Manchester M44 5BD United...

Page 10: ...ators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipment DATA ACQUISITION Data Acquisition Engineering Software Communications Based Acquisition Systems Plug in Cards for Apple IBM Compatibles Datalogging Systems Recorders Printers Plotter...

Reviews: