background image

OS100 SERIES

Mini-Infrared Transmitter

MADE IN

omega.com  

e-mail: [email protected]

For latest product manuals:

omegamanual.info

User’s Guide

Shop online at

OS101

OS102

Summary of Contents for OS101

Page 1: ...OS100 SERIES Mini Infrared Transmitter MADE IN omega com e mail info omega com For latest product manuals omegamanual info User sGuide Shop online at OS101 OS102 ...

Page 2: ...8280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056 9398 0 FAX 49 0 7056 9398 29 Toll Free in Germany 0800 639 7678 e mail info omega de United Kingdom One Omega Drive River Bend Technology Centre ISO 9002 Certified Northbank Irlam Manchester M44 5BD Un...

Page 3: ...on 3 1 3 1 Main Board 3 1 3 2 Ambient Temperature 3 2 3 3 Atmospheric Quality 3 3 3 4 Measuring Temperature 3 3 3 5 Alarm Setting 3 4 3 6 Adding Extension Cable 3 4 Section 4 Laser Sight Accessory 4 1 4 1 Warning and Cautions 4 1 4 2 Operating the Laser Sight Accessory 4 1 Section 5 Specifications 5 1 5 1 General 5 1 5 2 Laser Sight Accessory OS100 LS 5 2 Section 6 Emissivity Table 6 1 OS100 Serie...

Page 4: ...4 3 4 Setting the Temperature Engineering Unit 3 4 3 5 Mounting Bracket OS100 MB 3 5 3 6 Water Cooling Jacket OS100 WC 3 5 3 7 Typical Water Cool Jacket Assembly 3 5 3 8 Air Purge Collar OS100 AP 3 6 3 9 DIN Rail Mounting Adapter OS100 DR 3 6 3 10 NEMA 4 Aluminum Enclosure 3 6 4 1 Laser Sighting Accessory OS100 LS 4 2 4 2 Laser Warning Label 4 2 OS100 Series Infrared Transmitter Table of Figures ...

Page 5: ...ory is one 1 There are no user replaceable fuses in this product The output terminals of this product are for use with equipment digital meters chart recorders etc which have no accessible five parts Such equipment should comply with all the applicable safety requirements Do not operate the equipment in flammable or explosive environments All connections to the thermometer should be made via a shi...

Page 6: ...NOTES OS100 Series iv ...

Page 7: ...ia a 1 82 m 6 feet shielded cable as standard The unit provides field adjustable alarm output Section 2 Installation 2 1 Unpacking Remove the packing list and verify that you have received all your equipment If you have any questions about the shipment please call Customer Service at 1 800 622 2378 or 203 359 1660 We can also be reached on the internet omega com e mail info omega com When you rece...

Page 8: ...00 MB Mounting Bracket OS100 DR DIN Rail Mounting adapter OS100 WC Water Cooling jacket OS100 AP Air purge collar OS100 LS Laser Sighting OS100 CA15FT Sensor Head Extension Cable 4 6 m 15 OS100 CA25FT Sensor Head Extension Cable 7 6 m 25 TX8 100 8 Conductor stranded Shielded cable 30 m 100 PSU 93 Unregulated 16 24 VDC Power Supply CAL3 IR NIST Traceable Calibration Features OS101 OS102 Accuracy 2 ...

Page 9: ...Turn Potentiometer to adjust Emissivity in tenths 0 x_ 3 Single Turn Potentiometer to adjust Emissivity in hundreds 0 _x 4 Slide switch to select between real time Normal Operation and alarm set point 5 Alarm set point adjust P4 6 Sensor Head connection 7 Input Zero adjust P3 8 Input Span adjust P2 9 Output Zero adjust P5 10 Output Span adjust P6 ANALOG OUTPUTS K 4 20mA 1mV Deg 0 5VDC T C A V POWE...

Page 10: ...temperature measurement can be made When the ambient temperature around the sensor head changes abruptly the sensor head goes through thermal shock It takes a certain amount of time for the sensor head to stabilize to the new ambient temperature For example it takes about 30 minutes for the sensor head to stabilize going from 25 C to 50 C 77 to 122 F ambient temperature The sensor head dimensions ...

Page 11: ... slide switch SW1 on the main board is set to real time Fig 3 1 The target is larger than the optical field of view of the sensor head Fig 3 3 The emissivity adjustment on the main board is set properly Fig 3 1 The output load is within the product specification On OS102 transmitters follow these additional steps The temperature display is set to F or C Fig 3 4 For 4 20mA output models make sure a...

Page 12: ...t directly based on the temperature display 3 6 Adding Extension Cable You can add extension cable between the Sensor Head and the main electronic housing up to 15 2 m 50 feet After adding the extension cable the Zero input potentiometer P3 may be re adjusted See Fig 3 1 for proper analog output reading The following figures show the mounting bracket OS100 MB Water cooling jacket OS100 WC Air purg...

Page 13: ...00 Series Operation 3 23 3 919 Ø 19 0 755 Ø 4 0 156 31 8 1 25 90 3 2 1 8 12 7 0 5 12 7 0 5 50 0 1 97 3 2 1 8 25 4 1 42 9 1 69 Dimensions mm in INLET OUTLET 1 8 NPT COMPRESSION FITTING 2 PLCS 3 4 16 UNF 2B 3 4 16 UNF 41 3 1 625 8 312 60 3 2 375 Dimensions mm in INLET OUTLET TO MAIN HOUSING UP TO 200C 392F AMBIENT TEMPERATURE 4 3 2 1 ...

Page 14: ...OUSING 35 5 1 400 53 3 2 10 103 4 4 07 DIM A 78 0 3 07 47 7 1 880 5 0 0 20 5 7 0 225 DIA 4 HOLES 8 6 0 340 DIA x 22 6 0 890 DEEP 2 PLACES 65 5 2 578 115 4 4 545 Dimensions mm in DIM A OS101 26 4 1 04 OS102 51 8 2 04 44 5 1 75 109 5 4 3125 4 40 THRD HOLE 2 PLCS DIN RAIL SNAP FEET Dimensions mm in 28 6 1 125 1 16 NPT COMPRESSION FITTING 3 4 16 UNF 2B THRU 17 5 0 69 ...

Page 15: ...nside 4 2 Operating the Laser Sight Accessory The laser sight accessory screws onto the front of the sensor head This accessory is only used for alignment of the sensor head to the target area After the alignment process the accessory has to be removed from the front of the sensor head before temperature measurement The laser sight accessory is powered from a small compact battery pack included wi...

Page 16: ... CABLE SWITCH ON OFF BATTERY PACK POWER SUPPLY 50 8 2 0 19 75 38 1 1 5 Dimensions mm in OUTPUT 1 mW WAVELENGTH 630 670 nm CLASS II 2 LASER PRODUCT COMPLIES WITH FDA 21CFR 1040 10 EN60825 1 11 2001 LASER RADIATION DO NOT STARE INTO BEAM LASER RADIATION DO NOT STARE INTO BEAM OR VIEW DIRECTLY WITH OPTICAL INSTRUMENTS CLASS 2 LASER PRODUCT CAUTION AVOID EXPOSURE LASER RADIATION IS EMITTED FROM THIS A...

Page 17: ... OS100 WC Water Cooling Jacket 0 to 200 C 32 to 392 F Operating Relative Humidity Less than 95 RH non condensing Water Flow Rate for OS100 WC 0 25 GPM room temperature Thermal Shock About 30 minutes for 25 C abrupt ambient temperature change Warm Up Period 3 minutes Air Flow Rate for OS100 AP 1 CFM 0 5 Liters sec Power 12 to 24 VDC 100 mA Analog Outputs MV F 1 mV F MV C 1 mV C K K Type TC OS101 on...

Page 18: ...ent temperature European Classification Class 2 EN60825 1 11 2001 Maximum Operating current 45 mA at 3 VDC FDA Classification Complies with 21 CFR 1040 10 Class II Laser Product Beam Diameter 5 mm Beam Divergence 2 mrad Operating Temperature 0 to 50 C 32 to 122 F Operating Relative Humidity Less than 95 RH non condensing Power Switch ON OFF Slide switch on the Battery Pack Power Indicator Red LED ...

Page 19: ...dized dull wrought iron 0 94 Iron and steel excluding stainless rusted iron plate 0 69 Iron and steel excluding stainless polished steel 0 07 Iron and steel excluding stainless polished steel oxidized at 600 C 1112 F 0 79 Iron and steel excluding stainless rolled sheet steel 0 66 Iron and steel excluding stainless rough steel plate 0 94 to 0 97 Lead gray and oxidized 0 28 Mercury 0 09 to 0 12 Moly...

Page 20: ...yrex lead soda 0 85 to 0 95 Marble polished light gray 0 93 Paints lacquers and varnishes Black matte shellac 0 91 Paints lacquers and varnishes aluminum paints 0 27 to 0 67 Paints lacquers and varnishes flat black lacquer 0 96 to 0 98 Paints lacquers and varnishes white enamel varnish 0 91 Porcelain glazed 0 92 Quartz opaque 0 68 to 0 92 Roofing Paper 0 91 Tape Masking 0 95 Water 0 95 to 0 96 Woo...

Page 21: ...NOTES OS100 Series 6 3 ...

Page 22: ...NOTES OS100 Series 6 4 ...

Page 23: ...herwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans S...

Page 24: ...ity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipment DATA ACQUISITION Data Acquisition Engineering Software Communications Based Acquisition Systems Plug in Cards for Apple IBM Compatibles Datalogging Systems Recorders Printer...