DISTANCE AND SPOT SIZE

The size of the area (spot size) to be measured determines the distance between the sensor and 

the target.  The spot size must not be larger than the target.  The sensor should be mounted so 

that the measured spot size is smaller than the target.

AMBIENT TEMPERATURE

The sensor is designed to operate in ambient temperatures from 0°C to 75°C. 
Avoid thermal shock. Allow 20 minutes for the unit to adjust to large changes in ambient 

temperature.

ATMOSPHERIC QUALITY

Smoke, fumes or dust can contaminate the lens and cause errors in temperature measurement.  

In these types of environment the air purge collar should be used to help keep the lens clean. 

ELECTRICAL INTERFERENCE 

To minimise electromagnetic interference or ‘noise’, the sensor should be mounted away from 

motors, generators and such like.

MECHANICAL INSTALLATION

All sensors come with a 1.5 m USB cable and a mounting nut.  The sensor can be mounted on 

brackets, clamps or cut outs of your own design, or you can use the fixed and adjustable 

mounting bracket accessories which are shown below. 

 

AIR PURGE COLLAR

The air purge collar below is used to keep dust, fumes, moisture, and other contaminants away 

from the lens.  It must be screwed in fully.  Air flows into the 1/8” BSP fitting and out of the front 

aperture.  Air flow should be no more than 5 to 15 litres/min.

Clean or ‘instrument’ air is recommended.

12.0

45.0

50.0

50.0

40.0

9.0

15.0

25.0

25.0

60° Rotation

60° Rotation

60° Rotation

9.0

48.0

Fixed Bracket

Adjustable Bracket

2 x Mounting Holes M4 Clearance

2 x Mounting Holes M4 Clearance

9.0

24.0

24.0

Ø16.0

Summary of Contents for OS-MINIUSB series

Page 1: ...e mail info omega com For latest product manuals omegamanual info Shop online at omega comSM User sGuide OS MINIUSB SERIES...

Page 2: ...notice Servicing North America U S A Omega Engineering Inc One Omega Drive P O Box 4047 Stamford CT 06907 0047 USA Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada o...

Page 3: ...21 20 1 OS MINIUSB SN201 Interface USB Accuracy 1 of reading or 1 C whichever is greater Repeatability 0 5 of reading or 0 5 C whichever is greater Emissivity 0 2 to 1 0 Response Time t90 125 ms 90 re...

Page 4: ...collar Laser sighting tool OPTIONS An optional certificate of calibration is available for newly manufactured sensors If required it must be ordered at the same time as the sensor OPTICAL CHART The op...

Page 5: ...CE To minimise electromagnetic interference or noise the sensor should be mounted away from motors generators and such like MECHANICAL INSTALLATION All sensors come with a 1 5 m USB cable and a mounti...

Page 6: ...ntil the software has been fully installed The software will tell you when to connect the sensor This will ensure the USB driver for the sensor is installed properly 1 Insert the provided CD or downlo...

Page 7: ...ivity setting should match the actual emissivity of the target surface Non reflective non metals such as rubber foods thick plastics organic materials and painted surfaces generally have a high emissi...

Page 8: ...Status bits active high Bit 0 Measurement error Bit 1 Sensor temperature low Bit 2 Sensor temperature high Bit 3 Object temperature low Bit 4 Object temperature high R 0x16 1 Reflected Temperature no...

Page 9: ...with Windows Vista 7 8 or 10 Erroneous temperature Field of view obstruction Remove obstruction LENS CLEANING Keep the lens clean at all times Any foreign matter on the lens would affect measurement a...

Page 10: ...8...

Page 11: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 12: ...ity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: