background image

OM-USB-2404-60 User's Guide 

Specifications 

14 

Safety voltages 

Table 7. Safety specifications (Note 3) 

Parameter 

Conditions 

Specification 

Channel-to-earth ground isolation 

Continuous 

250 Vrms, Measurement Category II (Note 4) 

Withstand 

2,300 Vrms, verified by a 5 sec dielectric withstand test 

Channel-to-channel isolation 

Continuous 

250 Vrms, Measurement Category II (Note 4) 

Withstand 

1390 Vrms, verified by a 5 sec dielectric withstand test 

Note 3:  

Connect only voltages that are within the limits specified in this table. 

Note 4:  

Measurement Category II is for measurements performed on circuits directly connected to the 
electrical distribution system. This category refers to local-level electrical distribution, such as 
that provided by a standard wall outlet, for example 115 V for US or 230 V for Europe.  

Caution!

 

Do not connect the device to signals or use for measurements within Measurement Categories 
III or IV. 

Screw terminal connectors 

Table 8. Screw terminal connector specifications 

Connector type 

Screw terminal 

Screw terminal wiring 

16 to 28 AWG copper conductor wire with 7 mm (0.28 in.) of insulation stripped 
from the end. 

Torque for screw terminals 

0.22 to 0.25 N· m (1.95 to 2.21 lb. · in.) 

Table 9. Screw terminal assignments 

Screw terminal 

Signal 

CH0+ (CH0 IN HI) 

CH0

– (CH0 IN LO) 

CH1+ (CH1 IN HI) 

CH1

– (CH1 IN LO) 

CH2+ (CH2 IN HI) 

CH2

– (CH2 IN LO) 

CH3+ (CH3 IN HI) 

CH3

– (CH3 IN LO) 

Accessory products 

Table 10. Screw terminal connector specifications 

ACC-102 

Two-position detachable screw terminal connector blocks (quantity ten) 

ACC-160 

Backshell for use with the ACC-102 two-position screw terminal connector blocks. 
Provides strain relief and operator protection from high-voltage signals (quantity 
four) 

 

 

 

 

 

Summary of Contents for OM-USB-2404-60

Page 1: ......

Page 2: ......

Page 3: ...cessories 6 Unpacking 6 Installing the software 6 Installing the hardware 6 Calibrating the hardware 7 Chapter 3 Functional Details 8 Components 8 Screw terminals CH0 to CH3 8 USB connector 9 LED 9 St...

Page 4: ...able on our website at www omega com You can also contact Omega Engineering by phone fax or email with specific questions Phone 203 359 1660 Fax 203 359 7700 Email das omega com Safety guidelines You...

Page 5: ...sion A 250 Vrms channel to channel and channel to ground isolation protects the device and computer from ground spikes and ensures a reliable data stream An on board LED indicates the status of the de...

Page 6: ...touching the computer chassis or other grounded object to eliminate any stored static charge If any components are missing or damaged notify Omega Engineering immediately by phone fax or e mail Phone...

Page 7: ...e device This warm up time is required to achieve the specified rated accuracy of measurements Calibrating the hardware The OM USB 2404 60 is shipped fully calibrated Calibration coefficients are stor...

Page 8: ...for connection to four isolated analog inputs The maximum sampling rate is 50 kS s per channel and the input voltage range is 60 V nominal The channels may be sampled individually or simultaneously Si...

Page 9: ...peed host controller 480 Mbps and USB 2 0 high speed hubs Strain relief slot Use the strain relief slot to keep the USB cable from disconnecting from the device inadvertently Feed a tie wrap through t...

Page 10: ...for one analog input channel Figure 7 Analog input circuitry for one channel Filtering The OM USB 2404 60 uses a combination of analog and digital filtering to provide an accurate representation of i...

Page 11: ...artifact of signals at a higher frequency The alias free bandwidth 0 453 is defined by the ability of the filter to reject frequencies above the stopband frequency The alias free bandwidth equals the...

Page 12: ...mum Input voltage ranges CH to CH 60 V nominal 62 64 V typical 61 5 V minimum Overvoltage protection 100 V Input coupling DC Input impedance CH to CH 1 M Input noise 320 V rms Gain drift 5 ppm C Offse...

Page 13: ...er Table 3 Power specifications Parameter Specification Current consumption from USB 500 mA maximum Bus interface Table 4 Bus specifications Parameter Specification USB specification USB 2 0 Hi Speed...

Page 14: ...outlet for example 115 V for US or 230 V for Europe Caution Do not connect the device to signals or use for measurements within Measurement Categories III or IV Screw terminal connectors Table 8 Screw...

Page 15: ......

Page 16: ......

Reviews: