OM-NET-1608 User's Guide
Specifications
25
Counter
Table 13. Counter specifications
Parameter
Specification
Pin name
CTR
Counter type
Event counter
Number of channels
1
Input type
Schmitt trigger, 47 kΩ pull-down to ground
Input source
CTR screw terminal
Resolution
32 bits
Schmitt trigger hysteresis
1.01 V typ
0.6 V min
1.5 V max
Input high voltage threshold
2.43 V typ
1.9 V min
3.1 V max
Input high voltage limit
5.5 V absolute max
Input low voltage threshold
1.42 V typ
1.0 V min
2.0 V max
Input low voltage limit
–0.5 V absolute min
0 V recommended min
Input frequency
10 MHz max
High pulse width
50 ns min
Low pulse width
50 ns min
Memory
Table 14. Memory specifications
Parameter
Specification
Data FIFO (analog input)
49,152 samples
Non-volatile memory
2,048 bytes (768 bytes for calibration, 256 bytes for user, 1,024 bytes for network
settings)
Power
Table 15. Power specifications
Parameter
Condition
Specification
External power supply
5V, 1A
Supply current
Quiescent current
330 mA typical (Note 6)
710 mA max including all external loading
User output voltage range
Available at +VO terminal
4.40 V min to 5.25 V max, assumes supplied
AC adapter is used
User output current
Available at +VO terminal
10 mA max
Note 6:
This is the total quiescent current requirement for the device that includes the LEDs. This does not include any
potential loading of the digital I/O bits, +VO terminal, or the AOUTx outputs.