background image

OM-CP-RFCURRENT2000A

Product Overview

The OM-CP-RFCURRENT2000A is a wireless data logger with display, designed to measure and record DC  
current signals. The data logger can connect directly to low-level DC current sources which makes it ideal  
for battery usage studies, energy monitoring studies and other DC current monitoring applications. The  
OM-CP-RFCURRENT2000A can also be interfaced with a wide range of current output sensors such as flow  
meters, pH transmitters and more. Through the software, the device can be programmed with engineering 
units, which convert the current readings into a different unit of measure such as flow rate. The engineering 
unit is available for display on the LCD screen as well as in the software. The OM-CP-RFCURRENT2000A is 
available in 20mA, 160mA or a 3A measurement range.

The OM-CP-RFCURRENT2000A features an LCD screen which provides instant access to the current  
reading, minimum, maximum and average statistics. The statistics can be reset through the keypad of  
the device, allowing customers to keep track of statistics over certain time periods. This feature does not  
erase the data recorded, but only resets the statistics count. The OM-CP-RFCURRENT2000A can be used as a 
standalone data logger, or used with the OM-CP-RFC1000-EXT wireless transceiver, to transmit data back  
to a central PC. The OM-CP-RFCURRENT2000A transmits up to 500’ line of sight to the OM-CP-RFC1000-EXT 
transceiver. Multiple OM-CP-RFC1000-EXT’s can be used as repeaters and transmit up to 1000’ line of sight  
to other OM-CP-RFC1000-EXT’s. This allows the data to be sent over a greater distance.

Selection Buttons

The logger is designed with 3 direct selection buttons: Scroll, Units and Wireless.
• 

Scroll:

 Allows user to scroll through current readings, average stats and device status information displayed 

on the LCD Screen.

• 

Units: 

Allows user to change displayed units of measurement, A and mA Engineering Units.

• 

Wireless:

 Push and hold this button for 5 seconds to activate or deactivate wireless communication.

• 

Manual Reset Option:

 Allows user to manually reset the statistics within the device to 0 without needing to 

use the Omega software. Any data recorded up to that point is recorded and saved.

LED Indicators

• 

Status:

 Green LED blinks every 5 seconds to indicate unit is logging.

• 

Wireless:

 Blue LED blinks every 15 seconds to indicate unit is operating in wireless mode.

• 

Alarm:

 Red LED blinks every 1 second to indicate an alarm condition. 

Connecting Information
Port Information

The OM-CP-RFCURRENT2000A data logger terminal ports are located 
on the bottom of the device, one positive, one negative and a ground 
terminal port. 

GND

Summary of Contents for OM-CP-RFCURRENT2000A

Page 1: ...OM CP RFCURRENT2000A Wireless Current Data Logger...

Page 2: ...alone data logger or used with the OM CP RFC1000 EXT wireless transceiver to transmit data back to a central PC The OM CP RFCURRENT2000A transmits up to 500 line of sight to the OM CP RFC1000 EXT tran...

Page 3: ...data logger by clicking the Start icon 7 To download data select the device in the list click the Stop icon and then click the Download icon A graph will automatically display the data 1 Install the...

Page 4: ...fications will not be received if settings are incorrect 5 Click OK The OK button will be disabled if the Rule name box is empty Changing the Alarm Settings of a Device 1 In the Connected devices pane...

Page 5: ...on then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all of the available communication interfaces If the OM CP RFC1000 EXT is listed here then th...

Page 6: ...4 Push the cover closed until it clicks Recalibration Standard recalibration for the OM CP RFCURRENT2000A 3A are two points 0A and 1 00A For the OM CP RFCURRENT2000A 20mA the standard calibration poin...

Page 7: ...ature CO2 flow rate and more Calibration Digital calibration through software Calibration Date Automatically recorded within device Battery Type Ultralife 9V lithium battery included user replaceable...

Page 8: ...line of sight unobstructed 500 max indoors typical urban environment Measurement Wireless Battery Warning WARNING MAY LEAK FLAME OR EXPLODE IF DISASSEMBLED SHORTED CHARGED CONNECTED TOGETHER MIXED WI...

Page 9: ...Quick Start Manual OM CP RFC1000 EXT Wireless Transceiver Other models available OM CP RFC1000 EUR Approved for European markets OM CP RFC1000 IP69K Splash Proof for harsh environments...

Page 10: ...85 C and a humidity range of 0 to 95 RH non condensing The OM CP RFC1000 EXT is rated IP40 and is protected against solids that are greater than 1mm in size This device is not water resistant LEDs Th...

Page 11: ...performance both the OM CP RFC1000 EXT and the Omega data loggers should be mounted in the same orientation This usually means that the external antenna should be pointing straight up The antenna can...

Page 12: ...r The images below show the orientations available of the switches for each channel Channel 26 all switches in the up position is not supported Follow the instructions below to configure the channel s...

Page 13: ...o wireless mode by holding down the Wireless button for 5 seconds Deploying and Activating Devices 1 Plug the OM CP RFC1000 EXT into the USB port on the base station computer Additional OM CP RFC1000...

Page 14: ...OM CP RFC1000 EXT is not properly connected 1 In the software click the File button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all of the a...

Page 15: ...als FCC ID OA3MRF24J40MC IC 7693A 24J40MC Transmission Distance To other OM CP RFC1000 EXTs 4000 max outdoor line of sight unobstructed 1000 max indoors typical urban environment Transmission Distance...

Page 16: ...ustrie Canada applicables aux appareils radio exempts de licence L exploitation est autoris e aux deux conditions suivantes 1 l appareil ne doit pas produire de brouillage et 2 l utilisateur de l appa...

Page 17: ...l application used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA...

Reviews: