![Omega IP210A Series User Manual Download Page 6](http://html1.mh-extra.com/html/omega/ip210a-series/ip210a-series_user-manual_3987779006.webp)
M5736/0619 IP210A SERIES
A
6
TROUBLESHOOTING
Note:
If you experience an unusual problem that is not described in this section, or if troubleshooting procedures will not
restore the unit to normal operation, please contact Omega.
SYMPTOM
CAUSE
SOLUTION
1) Plugged or blocked orifice.
1) Remove orifice screw and clean
orifice with wire or pin <0.009"
diameter
2) Supply pressure too low.
2) Supply pressure must be at least 5
PSIG above the 100% output.
1) Zero set too low.
1) Turn zero adjustment counter-
clockwise until output pressure
increases then follow calibration
procedure.
2) Electrical connection has opened or
wrong type of signal applied.
2) Check electrical connections to
insure that there is a correct input
signal reaching the transducer.
Wires connected in reverse
Insure wiring is connected per Electrical
Connection section.
Pneumatic connections reversed or
supply and output both connected to
“IN” ports.
Insure supply is connected to port
marked “IN” and output is connected to
port marked “OUT”.
Sluggish Performance or Reduced
Range
matter the input signal3.
Transducer remains at start up no
signal goes up.)
(Pressure goes down when input
Transducer works in reverse.
Output Pressure same as supply
Pressure with no signal effect.
IP210A Unit shown without female 43650A connector.
Figure 3