background image

CALIBRATION

M5736/0619 IP210A SERIES

A

5

If 4 to 20 mA

0% = 4 mA

25% = 8 mA

50% = 12 mA

75% = 16 mA

100% = 20 mA

3.  Turn the “Span” adjusting screw counter-clockwise to increase output. 
4.  Repeat steps 1& 2 until unit is calibrated.

Reverse Acting

2.  Apply the 100% signal to unit and adjust “Span” adjusting screw until output is at the high setpoint.

Forward Acting

1.  Apply the 0% input signal to unit and adjust "Zero" adjusting screw until output reaches the high output pressure set-point. Turn 

counter-clockwise to increase.

2.  Apply the 100% input signal to unit and adjust "Span" adjustment until output pressure is at the low setpoint.

3.  Repeat steps 1& 2 until unit is calibrated.

IMPORTANT: Turn the “Span” adjusting screw clockwise to increase output

1.  Apply the 0% input signal to unit and adjust the “Zero” adjusting screw to the low setpoint output. Turn screw counter-clockwise to 

increase output. If there is no change in output, keep turning adjustment screw counter-clockwise until output begins to rise. It is 
imperative that the output pressure be closely monitored as turning the zero screw excessively could damage the flexure.

Note: input percentage refers to the following examples:

Once the transducer has been mounted, and air supply connections finished, the unit's calibration must be checked before placing into 
service. Remove rubber plugs to gain access to “ZERO” and “SPAN” adjusting screws. See Figure 3 for “ZERO” and “SPAN” locations. Re-
install rubber plugs after calibration is complete.

SIGNAL IMPEDANCE

INPUT SIGNAL

OUTPUT RANGE

NOMINAL IMPEDANCE

(OHMS)

POTENTIOMETER RV1 SIZE

(OHMS)

4-20mA 

3-15 PSIG 

180 

1000

4-20mA 

3-27 PSIG 

215 

1000

4-20mA 

3-120 PSIG 

260 

1000

4-20mA 

2-60 PSIG 

225 

1000

The input impedance will vary depending upon the potentiometer setting of the “SPAN (RV1)”, (

SEE FIG. 2

). The I/P circuit design consists 

of a resistor and potentiometer in parallel with the coil/flexure. The following are some of the nominal impedences that would be measured 
across the signal leads of the average calibrated transducer. These values are helpful when determining power requirements and for 
troubleshooting.

BLACK

WHITE

RV1

121 Ω

R2

100 Ω @ 25°C

COIL
300 Ω

J2

J1

I / P SCHEMATIC

RT1

R1

68 Ω

FIGURE 2

Summary of Contents for IP210A Series

Page 1: ...e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide IP210A SERIES ELECTROPNEUMATIC CONVERTERS...

Page 2: ...h America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 3...

Page 3: ...o rear of panel Attach transducer See Fig 3 to panel using two 10 32 screws and two threaded mountingholesonthebackoftheunit Panel with no access to rear of panel Attach bracket See Fig 1 and 3 to tra...

Page 4: ...the input wires entering the transducer is recommended should installationnexttohighelectromagneticinterferencebenecessary ToolsRequired Solder SolderingIron WireCutter Stripper Includedwiththeunit In...

Page 5: ...tivethattheoutputpressurebecloselymonitoredasturningthezeroscrewexcessivelycoulddamagetheflexure Note inputpercentagereferstothefollowingexamples Once the transducer has been mounted and air supply co...

Page 6: ...ow calibration procedure 2 Electrical connection has opened or wrong type of signal applied 2 Check electrical connections to insure that there is a correct input signal reaching the transducer Wires...

Page 7: ...moisture or vibration improper specification misapplication misuse or other operating conditions outside of OMEGA s control Components in which wear is not warranted include but are not limited to co...

Page 8: ...Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps I...

Reviews: