Frequently Asked Questions

FAQs

How often is a new pressure 

measurement displayed?

The calibrator display updates with a new pressure 
measurement five times per second.

What are the different Display functions 

on the calibrator?

The calibrator ships with five functions turned on. 
They are listed in bold below. Turn functions on or 
off using HHP400 Suite.

1. 

Home

 (default)

7.  AVG

 

(Average)

2. 

MIN

 

(Minimum)

8.  RATE

 

3. 

MAX

 

(Maximum)

9. 

DATA LOG

 

4.  RELIEF VALVE TEST

10. Temperature & Time 

5. 

+ / -

 (

Accuracy)

11. LEAK TEST 

6. 

T.OFF, T.ON

 

(Tare)

12. LEAK TEST 

DURATION

What does it mean when the red 

backlight flashes?

 

The flashing red backlight indicates an error 
condition.

 

Possible error conditions are:

 

Pressure has exceeded the stated upper 
limit of the sensor(s).

 

Pressure has fallen below the stated lower 
limit of the sensor(s).

 

The red backlight overrides the white backlight.

FAQs

What does the Zero (Ø) key do?

Home

 

If the sensor is within a tolerance band around 
zero, press and hold the 

Zero

 key to zero the 

pressure measurement and to reset the Min 
and Max measurements.

 

Note

: The tolerance band is approximately ± 1 

% of the Full Scale pressure value of the sensor.

 

Absolute sensors have additional functionality.

In Min or Max mode

 

Press and hold the 

Zero

 key to reset the Min 

and Max measurement. However, this does not 
zero the pressure measurement.

In Tare mode

 

When the Tare is off (T.OFF), press and hold 
the 

Zero

 key to turn on Tare (T.ON) and to 

set the Tare value at the current p ressure 
measurement.

 

Likewise, when the Tare is on (T.ON), press and 
hold the 

Zero

 key to turn off the Tare mode. 

In Average mode

 

Press and hold the 

Zero

 key to restart the 

rolling average.

Holding the Zero key

 

The key must be held to perform the Zero or 
Tare mode. The displayed value(s) dashes out 
during the zero or tare process. 

Why are the available units different 

between a 15 psi and 3 000 psi sensor?

If a given measurement unit cannot display 
the correct number of digits, the calibrator 
automatically advances to the next displayable unit.

FAQs

Which version of Data Log is on my 

calibrator?

1.  Turn on the calibrator.
2.  Press the Display button until 

DATA LOG

 

appears.

3.  Press the Information button. The version 

appears: 

Vers. Pro

 or 

Vers. Lite.

 

How long does the white backlight stay 

on after the Backlight key is pressed?

 

The white backlight has an automatic timeout. 
If no keys have been pressed, the backlight 
automatically turns off after 1 minute.

 

Note:

 You can configure the backlight timeout 

with HHP400 Suite.

What does the bar graph at the top of the 

display indicate?

 

The bar graph displays a live indication of the 
current pressure applied to the sensor as a % of 
Full Scale.

 

However, when the battery percentage is 
displayed, the bar graph follows the large digit 
display in indicating the remaining state of 
charge for the batteries.

What does low battery indicator mean?

 

Be prepared to change batteries when you 
see the outline of the battery icon or when the 
outline of the icon flashes.

FAQs

How long will the calibrator remain on if 

I leave it unattended?

 

The default setting is Always On. 

 

You can configure the timeout for Auto Off with 
HHP400 Suite.