![background image](http://html1.mh-extra.com/html/omega/hhp400/hhp400_user-manual_4173834056.webp)
M20011 REL-0000 (2103)
User Manual for Smart Manometer
58 of 60
Media compatibility
Pressure Types
▪
DN:
Differential pressure, non-isolated sensors for use with
clean, dry, non-corrosive gases only.
▪
CN:
Compound non-isolated sensors for use with clean, dry,
non-corrosive gases only.
▪
AN:
Absolute non-isolated sensors for use with clean, dry,
non-corrosive gases only.
▪
DI:
Differential pressure, isolated sensors for use with gases
and liquids compatible with 316LSS and Viton O-rings
▪
CI:
Compound or Absolute pressure sensors for use with
gases and liquids compatible with 316LSS.
▪
AI:
Compound or Absolute pressure sensors for use with
gases and liquids compatible with 316LSS.
Pressure Limits
▪
DN units:
3x range or 200 psi, whichever is less.
▪
CN units:
3x range or 200 psi, whichever is less.
▪
AN units:
3x range or 200 psi, whichever is less.
▪
DI units:
1000 psi common mode, P1 (HI) only is 3x range, P2
(LO) only is 3x range or 150 PSI, whichever is less.
▪
CI units:
2x range.
▪
AI units:
2x range.